Z-1203-2017Class II

The Schiller PB 1000, also known as: Merge Hemo Patient Data Module (PDM), V2 model, a major component of the...

Merge Healthcare Incorporated / initiated 2015-05-31 / Terminated / root cause: manufacturing process

Merge Healthcare, Inc. began a recall of The Schiller PB 1000, also known as: Merge Hemo Patient Data Module (PDM), V2 model, a major component of the Merge Hemo software on . FDA classified it as Class II and gives the reason as "Non-invasive blood pressure pumps (NIBP) voltage is not sufficient and causes the NIBP pumps to fail prior to reaching adequate initial pressure". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1203-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merge Healthcare Incorporated
Product
The Schiller PB 1000, also known as: Merge Hemo Patient Data Module (PDM), V2 model, a major component of the Merge Hemo software.
Product code
DQK Computer, Diagnostic, Programmable
Reason for recall
Non-invasive blood pressure pumps (NIBP) voltage is not sufficient and causes the NIBP pumps to fail prior to reaching adequate initial pressure.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The recalling firm issued letters via email dated 3/28/2016 and 7/11/2016, or via certified mail if no email address was available. A third letter dated 1/23/2017 was also issued..
Quantity in commerce
341 sites potentially have the affected product versions
Distribution
Distribution was nationwide to medical facilities. There was also military and government distribution.
Product codes and lots
Patient Data Module (PDM) V2 model: Serial numbers 10.03.XXXXX

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1203-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1203-2017Class IITerminatedVoluntary: Firm initiated