Z-0599-2017Class II

Merge PACS software

Merge Healthcare Incorporated / initiated 2016-03-28 / Terminated / root cause: software

Merge Healthcare, Inc. began a recall of Merge PACS software. Product Usage: Merge PACS (Picture Archiving Communication System) is designed and marketed for soft copy reading, communication and... on . FDA classified it as Class II and gives the reason as "The software did not show unviewed images when the last view was skipped. The physician may potentially misdiagnose when not all images are available for viewing". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0599-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merge Healthcare Incorporated
Product
Merge PACS software. Product Usage: Merge PACS (Picture Archiving Communication System) is designed and marketed for soft copy reading, communication and storage of studies produced by digital modalities, including digital mammography.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
The software did not show unviewed images when the last view was skipped. The physician may potentially misdiagnose when not all images are available for viewing.
Root cause tag
software
FDA root cause
Software design
Action
The recalling firm issued a letter dated March 28, 2016 via e-mail or via certified mail to customers. The letter identified the affected product, problem and actions to be taken. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. The notification also informs the customer their response is required using the enclosed form and the return addressed envelope.
Quantity in commerce
534 sites potentially have the affected versions
Distribution
Worldwide Distribution - US Nationwide including Canada.
Product codes and lots
Versions: V6.0 MR2, 6.0 MR3, 6.0 MR3 CU, 6.0 MR4, 6.0 MR4 CU1, 6.0 MR4 CU2, 6.0 MR4 CU3, 6.0 MR4 CU4, 6.0 MR4 CU5, 6.0 MR4 CU6, 6.0 MR4, CU7, 6.0 MR4 CU8, 6.0 MR4 CU9, 6.0 MR4 CU10, 6.0 MR4 CU11, 6.1, 6.1.1, 6.1.2, 6.1.3, 6.2, 6.2.1, 6.2.2, 6.2.3, 6.2.4, 6.2.5, 6.3, 6.4, 6.4.1, 6.4.2, 6.4.3, 6.4.4, 6.4.5, 6.4.6, 6.4.7, 6.5, 6.5.5, 6.5.6, 6.5.7, 6.5.8, 6.5.9, 6.6, 6.6.1, 6.6.1.1, 6.6.2, 6.6.2.1, 6.6.2.2, 7.0, 7.0.1, and 7.0.2.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0599-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0599-2017Class IITerminatedVoluntary: Firm initiated