Merge PACS software
Merge Healthcare, Inc. began a recall of Merge PACS software. Product Usage: Merge PACS (Picture Archiving Communication System) is designed and marketed for soft copy reading, communication and... on . FDA classified it as Class II and gives the reason as "The software did not show unviewed images when the last view was skipped. The physician may potentially misdiagnose when not all images are available for viewing". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0599-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merge Healthcare Incorporated
- Product
- Merge PACS software. Product Usage: Merge PACS (Picture Archiving Communication System) is designed and marketed for soft copy reading, communication and storage of studies produced by digital modalities, including digital mammography.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- The software did not show unviewed images when the last view was skipped. The physician may potentially misdiagnose when not all images are available for viewing.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The recalling firm issued a letter dated March 28, 2016 via e-mail or via certified mail to customers. The letter identified the affected product, problem and actions to be taken. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. The notification also informs the customer their response is required using the enclosed form and the return addressed envelope.
- Quantity in commerce
- 534 sites potentially have the affected versions
- Distribution
- Worldwide Distribution - US Nationwide including Canada.
- Product codes and lots
- Versions: V6.0 MR2, 6.0 MR3, 6.0 MR3 CU, 6.0 MR4, 6.0 MR4 CU1, 6.0 MR4 CU2, 6.0 MR4 CU3, 6.0 MR4 CU4, 6.0 MR4 CU5, 6.0 MR4 CU6, 6.0 MR4, CU7, 6.0 MR4 CU8, 6.0 MR4 CU9, 6.0 MR4 CU10, 6.0 MR4 CU11, 6.1, 6.1.1, 6.1.2, 6.1.3, 6.2, 6.2.1, 6.2.2, 6.2.3, 6.2.4, 6.2.5, 6.3, 6.4, 6.4.1, 6.4.2, 6.4.3, 6.4.4, 6.4.5, 6.4.6, 6.4.7, 6.5, 6.5.5, 6.5.6, 6.5.7, 6.5.8, 6.5.9, 6.6, 6.6.1, 6.6.1.1, 6.6.2, 6.6.2.1, 6.6.2.2, 7.0, 7.0.1, and 7.0.2.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0599-2017 | Class II | Terminated | Voluntary: Firm initiated |