Z-0723-2017Class II

ADVIA 560 Hematology Systems, Siemens Material Number (SMN) 11170842, IVD

Siemens Healthcare Diagnostics Inc. / initiated 2016-10-21 / Terminated / root cause: software

Siemens Healthcare Diagnostics, Inc. began a recall of ADVIA 560 Hematology Systems, Siemens Material Number (SMN) 11170842, IVD on . FDA classified it as Class II and gives the reason as "Siemens is investigating an issue which may cause an incorrect result to be reported. Siemens received two reports of multiple discordant records for the same Sample ID in the ADVIA 560 Hematology System database that...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0723-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
ADVIA 560 Hematology Systems, Siemens Material Number (SMN) 11170842, IVD.
Product code
GKZ Counter, Differential Cell
Reason for recall
Siemens is investigating an issue which may cause an incorrect result to be reported. Siemens received two reports of multiple discordant records for the same Sample ID in the ADVIA 560 Hematology System database that occurred during the installation of the systems. The database should only contain one record of a Sample ID number for any given time and date. If there are multiple records for the same Sample ID, it is possible that multiple results may be manually or automatically sent to the Laboratory Information System (LIS), printed or displayed on the results report screen.
Root cause tag
software
FDA root cause
Software design
Action
An Urgent Medical Device Correction Letter HI17-02.A.US (dated October 21, 2016) and Response Form were sent to US Customers via Federal Express. An Urgent Field Safety Notice was sent to customers outside the US on October 21, 2016, (distribution is determined at the country level). The letters instruct customers to contact their local Siemens technical support representative if they obtain multiple results for one Sample ID. It also informs the customer what to do if the ADVIA 560 Hematology System generates any of the six error messages listed in the letter. For further questions, please call (312) 275-7795.
Quantity in commerce
US: 23 systems; Foreign: 141 systems
Distribution
Worldwide Distribution - US Distribution to the states of including Puerto Rico and : CA, CO, FL, IL, MI, NJ, NY, OH, TX, WA, WI, WI and WY., and to the countries of : Angola, Australia, Belgium, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Egypt, France, Germany, Greece, India, Italy, Japan, Kenya, Malaysia, Mexico, Myanmar, Netherlands, Norway, Pakistan, Portugal, Saudi Arabia, Slovenia, Spain, Switzerland, Turkey, Uganda, United Kingdom, Vietnam.
Product codes and lots
All serial numbers of ADVIA 560 Hematology Systems

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0723-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0723-2017Class IITerminatedVoluntary: Firm initiated