510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0724-2017Class II

Halyard C-Section Drape with Clear Screen and Full Incise (Non-Sterile), Product Code 44965 NS; C-Section...

Halyard Health / initiated 2016-11-09 / Terminated / root cause: manufacturing process

Halyard Health, Inc began a recall of Halyard C-Section Drape with Clear Screen and Full Incise (Non-Sterile), Product Code 44965 NS; C-Section Drape with Clear Screen and Full Incise (Sterile)... on . FDA classified it as Class II and gives the reason as "Drapes within the affected lots may include a manufacturing variation that prevents convenient removal of the coated release paper, which inhibits easy access to the surgical site". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0724-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Halyard Health
Product
Halyard C-Section Drape with Clear Screen and Full Incise (Non-Sterile), Product Code 44965 NS; C-Section Drape with Clear Screen and Full Incise (Sterile), Product Code 44966 00; C-Section Drape with Clear Screen (Non-Sterile), Product Code 44967 NS; C-Section Drape with Clear Screen (Sterile), Product Code 44968 00; C-Section Drape with Pouch and Fenestration (Non-Sterile), Product Code 44977 NS; C-Section Drape with Pouch and Fenestration (Sterile), Product Code 44978 00 Product Usage: A surgical drape is a device made of natural or synthetic materials intended to be use as a protective patient covering, such as to isolate a site or surgical incision from microbial and other contamination.
Product code
KKX Drape, Surgical, Antimicrobial
Reason for recall
Drapes within the affected lots may include a manufacturing variation that prevents convenient removal of the coated release paper, which inhibits easy access to the surgical site.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Accounts were notified of the recall by letter on 11/9/2016. The consignees were instructed to quarantine and discontinue use of the recalled lots and to discard/destroy per the facility's standard disposal procedure for nonwoven materials. A response form was included which was to be sent back.
Quantity in commerce
132 cases
Distribution
US Nationwide Distribution in the states of AL, FL, GA, IL, KS, MN, MO, NE, NJ, OH, PH, TN, VA, WI, WY
Product codes and lots
Product Code 44965 NS - Lot Numbers AC6174###, AC62266###, AC6229###, AC6243###, AC6252###; Product Code 44966 00 - Lot Numbers AC6146##L, AC6157##L, AC6177##L, AC6215##L, AC6221##L, AC6228##L, AC6245##L, AC6252##L, AC6258##L, AC6272##L, AC6277##L; Product Code 44967 NS - Lot Numbers AC6170###, AC6219###, AC6223###, AC6229###, AC625###; Product Code 44068 00 - Lot Numbers AC6146##L, AC6177##L, AC6230##L, AC6231##L, AC6234##L, AC6242##L, AC6245##L, AC6264##L, AC6271##L, AC6277##L; Product Code 44077 NS - Lot Numbers AC6170###, AC6213###, AC6214###, AC6222###, AC6223###, AC6251###; Product Code 44978 00 - Lot Numbers AC6147##L, AC6177##L, AC6243##L, AC 6252##L - (# = includes numbers 0, 1, 2, 3, 4, 5, 6, 7, 8 or 9 - L = includes letters a, b, or c)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0724-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0724-2017Class IITerminatedVoluntary: Firm initiated