Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1092-2017Class II

Tina-Quant (Latex) HS Test System (C-Reactive Protein (Latex) High Sensitive) The Tina-Quant CRP (Latex) High...

Roche Diagnostics GmbH / initiated 2016-02-26 / Terminated

Roche Diagnostics Operations, Inc. began a recall of Tina-Quant (Latex) HS Test System (C-Reactive Protein (Latex) High Sensitive) The Tina-Quant CRP (Latex) High Sensitive Immunoturbidimetric assay is for the... on . FDA classified it as Class II and gives the reason as "Tina-quant Cardiac C-reactive Protein (Latex) High Sensitive (CRPHS) lot number 604450 (catalog number 11972855216) shows an under-recovery of up to -25% at concentrations below 5 mg/L on the MODULAR ANALYTICS P module...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1092-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Tina-Quant (Latex) HS Test System (C-Reactive Protein (Latex) High Sensitive) The Tina-Quant CRP (Latex) High Sensitive Immunoturbidimetric assay is for the in-vitro quantitative determination of C-reactive protein (CRP) in human serum and plasma on Roche automated clinical chemistry analyzers. Highly sensitive measurement of CRP is of use for the detection and evaluation of inflammatory disorders, and associated diseases, infection and tissue injury. Measurement of CRP may also be used as an aid in the assessment of the risk of future coronary heart disease. When used as an adjunct to other laboratory evaluation methods of acute coronary syndromes, it may also be an additional independent indicator of recurrent event prognosis in patients with stable coronary disease or acute coronary syndrome
Product code
NQD Cardiac C-Reactive Protein, Antigen, Antiserum, And Control
Reason for recall
Tina-quant Cardiac C-reactive Protein (Latex) High Sensitive (CRPHS) lot number 604450 (catalog number 11972855216) shows an under-recovery of up to -25% at concentrations below 5 mg/L on the MODULAR ANALYTICS P module. this issue can lead to erroneous low CRPHS results. CRPHS is used as a marker for the prediction of cardiovascular risk. An erroneous low result could lead to an incorrect cardiovascular risk assessment and subsequent delay of further examination.
FDA root cause
Under Investigation by firm
Action
Roche Diagnostics sent a Reagent Bulletin dated February 26, 2016, to all affected consignees. The recall notifications included a description of the reason for the recall, affected product, request for replacement, and consignee responsibilities; the notification did not include instructions for responding to the recall notification. Actions Required " Do not use CRPHS lot number 604450. If you have any affected product in your inventory, discard it per local guidelines. " Complete the attached replacement request form to receive replacement product. " File this Reagent Bulletin for future reference. Questions Please contact the Roche Support Network Customer Support Center at 1-800-428-2336 if you have questions about the information contained in this Reagent Bulletin.
Quantity in commerce
3204
Distribution
Nationwide Distribution including AL, AZ, CA, CT, FL, GA, KS, MA, MO, NJ, NY, PA, and WA.
Product codes and lots
Catalog number: 11972855216, lot number 60445001

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1092-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1092-2017Class IITerminatedVoluntary: Firm initiated