Z-0726-2017Class II

Merge PACS software. Merge PACS (Picture Archiving Communication System) is designed and marketed for soft...

Merge Healthcare Incorporated / initiated 2016-03-28 / Terminated / root cause: design

Merge Healthcare, Inc. began a recall of Merge PACS software. Merge PACS (Picture Archiving Communication System) is designed and marketed for soft copy reading, communication and storage of studies... on . FDA classified it as Class II and gives the reason as "Cut lines on the image may present horizontally rather than vertically". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0726-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merge Healthcare Incorporated
Product
Merge PACS software. Merge PACS (Picture Archiving Communication System) is designed and marketed for soft copy reading, communication and storage of studies produced by digital modalities, including digital mammography.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Cut lines on the image may present horizontally rather than vertically.
Root cause tag
design
FDA root cause
Component design/selection
Action
MERGE sent an Urgent Medical Device Recall dated March 28, 2016, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Merge Healthcare is working on correcting the issues and will inform customers once a fix is made available. Customers were asked to reply using the enclosed form and the return addressed envelope. If customers have further distributed the affected product they need to identify their customers and notify them at once of the product recall. Customers with questions were instructed to send an email to [email protected].
Quantity in commerce
537 sites potentially have the affected versions
Distribution
Worldwide Distribution - US (nationwide) Canada and foreign distribution
Product codes and lots
Versions V6 GA, 6.0 MR1, V6 MR1 CU1,V6 MR1 CU1 SF1, V6 MR1 CU4, V6 MR1 CU5, V6 MR1 CU6, V6 MR1 CU7,V6 MR1 CU8, 6.0 MR2, 6.0 MR3, 6.0 MR3 CU, 6.0 MR4, 6.0 MR4 CU1, 6.0 MR4 CU2, 6.0 MR4 CU3, 6.0 MR4 CU4, 6.0 MR4 CU5, 6.0 MR4 CU6, 6.0 MR4, CU7, 6.0 MR4 CU8, 6.0 MR4 CU9, 6.0 MR4 CU10, 6.0 MR4 CU11, 6.1, 6.1.1, 6.1.2, 6.1.3, 6.2, 6.2.1, 6.2.2, 6.2.3, 6.2.4, 6.2.5, 6.3, 6.4, 6.4.1, 6.4.2, 6.4.3, 6.4.4, 6.4.5, 6.4.6, 6.4.7, 6.5, 6.5.5, 6.5.6, 6.5.7, 6.5.8, 6.5.9, 6.6, 6.6.1, 6.6.1.1, 6.6.2, 6.6.2.1, 6.6.2.2, 7.0, 7.0.1, and 7.0.2

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0726-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0726-2017Class IITerminatedVoluntary: Firm initiated