Z-1207-2017Class II

Maquet Oasis Dry Suction Water Seal Chest Drain, Single Collection w/AC Reference: 3600-100

Atrium Medical Corporation / initiated 2016-12-23 / Terminated / root cause: manufacturing process

Atrium Medical Corporation began a recall of Maquet Oasis Dry Suction Water Seal Chest Drain, Single Collection w/AC Reference: 3600-100 Intended Use: To evacuate air and/or fluid from the chest cavity or... on . FDA classified it as Class II and gives the reason as "Chest drains manual vent valve located at the top of affected chest drains could potentially not re-seal after being depressed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1207-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Atrium Medical Corporation
Product
Maquet Oasis Dry Suction Water Seal Chest Drain, Single Collection w/AC Reference: 3600-100 Intended Use: To evacuate air and/or fluid from the chest cavity or mediastinum. " To help re-establish lung expansion and restore breathing dynamics. " To facilitate postoperative collection and reinfusion of autologous blood from the patients pleural cavity or mediastinal area
Product code
CAC Apparatus, Autotransfusion
Reason for recall
Chest drains manual vent valve located at the top of affected chest drains could potentially not re-seal after being depressed.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Maquet issued a Field action correction notification sent to customers on 12/23/16 via Fed'x for delivery on Tuesday Dec.27, 2016.who have received affected product and describes the failure mode. Accounts to complete the enclosed Field Notification Reply Form to acknowledge receipt of the notification. Return the Field Notification Reply Form to the following e-mail address: [email protected] or you may fax the form to 1-603-386-6590. Questions, contact your local Atrium/Maquet/Getinge chest drainage representative or Atrium Customer Service at 1-800-528-7486 (603-880-1433), Monday through Friday between 8:00 am and 5:00 pm (EST). UPDATE: On 1/16/17, Maquet issued recall letters that identified correct lot numbers and products in the letter titles: UPDATE-UPDATED AFFECTED DEVICES. Two letters issued one to customers who received the Dec 23 letter and one to new customers who who did not receive the Dec. 23 letter.
Quantity in commerce
2933 cases (17598 units)
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Canada and Mexico
Product codes and lots
Lot Number: 242909, 242910, 242912, 242914, 242915

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1207-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1205-2017Class IITerminatedVoluntary: Firm initiated
Z-1206-2017Class IITerminatedVoluntary: Firm initiated
Z-1207-2017Class IITerminatedVoluntary: Firm initiated
Z-1208-2017Class IITerminatedVoluntary: Firm initiated
Z-1209-2017Class IITerminatedVoluntary: Firm initiated
Z-1210-2017Class IITerminatedVoluntary: Firm initiated
Z-1211-2017Class IITerminatedVoluntary: Firm initiated