Maquet Express Dry Seal Suction Chest Drain, Single Collection w/AC Reference: 4000-100N
Atrium Medical Corporation began a recall of Maquet Express Dry Seal Suction Chest Drain, Single Collection w/AC Reference: 4000-100N Intended Use: To evacuate air and/or fluid from the chest cavity or... on . FDA classified it as Class II and gives the reason as "Chest drains manual vent valve located at the top of affected chest drains could potentially not re-seal after being depressed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1208-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Atrium Medical Corporation
- Product
- Maquet Express Dry Seal Suction Chest Drain, Single Collection w/AC Reference: 4000-100N Intended Use: To evacuate air and/or fluid from the chest cavity or mediastinum. " To help re-establish lung expansion and restore breathing dynamics. " To facilitate postoperative collection and reinfusion of autologous blood from the patients pleural cavity or mediastinal area
- Product code
- CAC Apparatus, Autotransfusion
- Reason for recall
- Chest drains manual vent valve located at the top of affected chest drains could potentially not re-seal after being depressed.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Maquet issued a Field action correction notification sent to customers on 12/23/16 via Fed'x for delivery on Tuesday Dec.27, 2016.who have received affected product and describes the failure mode. Accounts to complete the enclosed Field Notification Reply Form to acknowledge receipt of the notification. Return the Field Notification Reply Form to the following e-mail address: [email protected] or you may fax the form to 1-603-386-6590. Questions, contact your local Atrium/Maquet/Getinge chest drainage representative or Atrium Customer Service at 1-800-528-7486 (603-880-1433), Monday through Friday between 8:00 am and 5:00 pm (EST). UPDATE: On 1/16/17, Maquet issued recall letters that identified correct lot numbers and products in the letter titles: UPDATE-UPDATED AFFECTED DEVICES. Two letters issued one to customers who received the Dec 23 letter and one to new customers who who did not receive the Dec. 23 letter.
- Quantity in commerce
- 681 cases (4086 units)
- Distribution
- Worldwide Distribution - US (nationwide) and Internationally to Canada and Mexico
- Product codes and lots
- Lot Number: 242919, 242920
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1205-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1206-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1207-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1208-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1209-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1210-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1211-2017 | Class II | Terminated | Voluntary: Firm initiated |