510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0711-2017Class II

Empowr Knee System

Encore Medical, LP / initiated 2016-11-15 / Terminated / root cause: labeling and UDI

Encore Medical, Lp began a recall of Empowr Knee System Product Usage: Non-porous proximal tibial baseplate implants, utilized in either posterior cruciate ligament (PCL) retaining or PCL... on . FDA classified it as Class II and gives the reason as "A size 4 Empowr baseplate was mislabeled as a size 6. The affected lot was laser marked as M6, marked with the lot number for the size 6 minus lot, and distributed as size 6 minus". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0711-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encore Medical, LP
Product
Empowr Knee System Product Usage: Non-porous proximal tibial baseplate implants, utilized in either posterior cruciate ligament (PCL) retaining or PCL substituting total knee arthroplasty procedure.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
A size 4 Empowr baseplate was mislabeled as a size 6. The affected lot was laser marked as M6, marked with the lot number for the size 6 minus lot, and distributed as size 6 minus.
Root cause tag
labeling and UDI
FDA root cause
Mixed-up of materials/components
Action
The recalling firm sent an Urgent Field Safety Notice letter dated November 15, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions call 760-734-3551.
Quantity in commerce
7 units
Distribution
US Nationwide Distribution in the states of CO, IL, IN, FL, CA, AZ
Product codes and lots
033T1022

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0711-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0711-2017Class IITerminatedVoluntary: Firm initiated