Z-0721-2017Class II

RayStation 4.0, 4.5, 4.7 and 5.0; Radiation Therapy Treatment Planning System

RaySearch Laboratories AB (publ) / initiated 2016-11-09 / Terminated / root cause: software

RAYSEARCH LABORATORIES AB began a recall of RayStation 4.0, 4.5, 4.7 and 5.0; Radiation Therapy Treatment Planning System Product Usage: RayStation is a software system designed for treatment planning... on . FDA classified it as Class II and gives the reason as "Software anomaly; an issue was found with the proton Pencil Beam Scanning (PBS) dose calculation in RayStation 4.0, 4.5, 4.7 and 5.0. For treatment plans with a combination of range shifter, large air gap and beams that...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0721-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
RaySearch Laboratories AB (publ)
Product
RayStation 4.0, 4.5, 4.7 and 5.0; Radiation Therapy Treatment Planning System Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
Software anomaly; an issue was found with the proton Pencil Beam Scanning (PBS) dose calculation in RayStation 4.0, 4.5, 4.7 and 5.0. For treatment plans with a combination of range shifter, large air gap and beams that enter the patient surface at an oblique angle, the dose calculation accuracy may be less than expected.
Root cause tag
software
FDA root cause
Software design
Action
Ray Search sent an Urgent Field Safety Notice Medical Device Correction letter (dated 11/8/2016) via e-mailed to customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to Inspect their product and identify all installed units with the above software version number, then confirm you have read and understood this notice (contact information included).
Quantity in commerce
US: 66 units
Distribution
US Nationwide Distribution in the states of California, Florida, Illinois, Louisiana, Michigan, New Jersey, Tennessee, Texas, Washington
Product codes and lots
Build numbers: 4.0.2.9, 4.0.3.4, 4.5.0.19, 4.5.1.14, 4.5.2.7, 4.7.0.15, 4.7.1.10, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4, 5.0.0.37, 5.0.1.11 or 5.0.2.35

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0721-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0721-2017Class IITerminatedVoluntary: Firm initiated