Z-0789-2017Class II

Acuvance2 Safety Catheter, Product No. 1720-AI

Abbott Medical / initiated 2016-11-21 / Terminated / root cause: manufacturing process

Smiths Medical ASD Inc. began a recall of Acuvance2 Safety Catheter, Product No. 1720-AI Product Usage: The Acuvance 2 Safety Catheter is a Peripheral Intravenous Catheter (PIVC). PIVCs are designed... on . FDA classified it as Class II and gives the reason as "Smiths Medical became aware that three (3) lots of Acuvance2 Safety Catheter, product number 1720-AI, contain devices with missing or incorrectly positioned side port valves". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0789-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Acuvance2 Safety Catheter, Product No. 1720-AI Product Usage: The Acuvance 2 Safety Catheter is a Peripheral Intravenous Catheter (PIVC). PIVCs are designed for single use venipuncture on all patient populations where access to the veins and arteries are required. The catheter has a Safety Needle Assembly (SAN) designed to reduce risk of accidental needle sticks to the operators. The lateral port is equipped with a non-return silicone valve and a polypropylene snap cap with a winged polypropylene hub for fastening.
Product code
FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Reason for recall
Smiths Medical became aware that three (3) lots of Acuvance2 Safety Catheter, product number 1720-AI, contain devices with missing or incorrectly positioned side port valves.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Consignees were sent on 11/21/2016 a Smiths Medical "Urgent Medical Device Field Safety Notice: dated November 18, 2016. The letter was addressed to "Clinician who uses the Acuvance2 Safety Catheter". The letter described the Reason for Recall, Risk to Health, and Instructions to Customers. Requested consignees to complete and return the "Urgent Medical Device Field Safety Notice Response Form" within 10 days of receipt. For questions contact Stericycle via email at the [email protected].
Quantity in commerce
26,750
Distribution
Foreign Distribution in the Italy, France, Switzerland, Germany, Sweden
Product codes and lots
3120882, 3120883, 3220733

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0789-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0789-2017Class IITerminatedVoluntary: Firm initiated