Z-0761-2017Class II

The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in outdoor and indoor...

Physio Control, Inc. / initiated 2010-03-11 / Terminated / root cause: manufacturing process

Physio-Control, Inc. began a recall of The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in outdoor and indoor emergency care settings within the environmental... on . FDA classified it as Class II and gives the reason as "Customer complaint was received relating to a LifePak15 device that would power on then power down unexpectedly after a short period of use. When the LP15 powered down, power could be immediately restored by activation...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0761-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Physio Control, Inc.
Product
The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in outdoor and indoor emergency care settings within the environmental conditions specified. Product Part Numbers V-15-2-000XXX
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Customer complaint was received relating to a LifePak15 device that would power on then power down unexpectedly after a short period of use. When the LP15 powered down, power could be immediately restored by activation of the power button. The device was connected to a cellular modem from MultiTech systems for use with LIFE NET services. The corrective action was completed by Velocitor Solutions in December 2010.
Root cause tag
manufacturing process
FDA root cause
Environmental control
Action
Velocitor initiated this field action and notified affected customers on March 11, 2010. This action was reported to Physio-Control as completed on 12/21/2010.
Quantity in commerce
1399 units (866 in the US and 533 outside US)
Distribution
worldwide
Product codes and lots
This action was taken by Velocitor Solutions to notify their customers in 2010. The consignee list was not provided to Physio-Control.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0761-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0761-2017Class IITerminatedVoluntary: Firm initiated