The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in outdoor and indoor...
Physio-Control, Inc. began a recall of The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in outdoor and indoor emergency care settings within the environmental... on . FDA classified it as Class II and gives the reason as "Customer complaint was received relating to a LifePak15 device that would power on then power down unexpectedly after a short period of use. When the LP15 powered down, power could be immediately restored by activation...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0761-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Physio Control, Inc.
- Product
- The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in outdoor and indoor emergency care settings within the environmental conditions specified. Product Part Numbers V-15-2-000XXX
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- Customer complaint was received relating to a LifePak15 device that would power on then power down unexpectedly after a short period of use. When the LP15 powered down, power could be immediately restored by activation of the power button. The device was connected to a cellular modem from MultiTech systems for use with LIFE NET services. The corrective action was completed by Velocitor Solutions in December 2010.
- Root cause tag
- manufacturing process
- FDA root cause
- Environmental control
- Action
- Velocitor initiated this field action and notified affected customers on March 11, 2010. This action was reported to Physio-Control as completed on 12/21/2010.
- Quantity in commerce
- 1399 units (866 in the US and 533 outside US)
- Distribution
- worldwide
- Product codes and lots
- This action was taken by Velocitor Solutions to notify their customers in 2010. The consignee list was not provided to Physio-Control.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0761-2017 | Class II | Terminated | Voluntary: Firm initiated |