IQon Spectral CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body...
Philips Medical Systems (Cleveland) Inc began a recall of IQon Spectral CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission... on . FDA classified it as Class II and gives the reason as "A software defect is causing issues with the IQon Spectral CT and Philips Spectral CT Applications while filming, saving and annotating images". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0705-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- IQon Spectral CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- A software defect is causing issues with the IQon Spectral CT and Philips Spectral CT Applications while filming, saving and annotating images
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- A Field Safety Notice (Attachment #1 Field Safety Notice FSN 16-061-72800659 and FSN 88100051), will be sent to US consignees via certified mail. Philips Key Market representatives will distribute the notice internationally, based on local regulations. Additionally, a Field Service Engineer will contact each site to schedule a time to implement the Field Change Order.
- Quantity in commerce
- 24
- Distribution
- Software was distributed in the following states: AR, IN, LA, MA, MN, OR and TN. Softward was also shipped to the following countries: Belgium, Denmark, France, Germany, Japan, Korea, Switzerland and United Kingdom.
- Product codes and lots
- 860008 860009 860010 860011 860012 860013 860014 860015 860016 860017 860018 860019 860020 860021 860022 860023 860024 860028 860026 860029 860030 860031 860034 860035
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0705-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0706-2017 | Class II | Terminated | Voluntary: Firm initiated |