Z-0706-2017Class II

Philips Spectral CT Applications supports viewing and analysis of images at energies selected from the...

Philips North America LLC / initiated 2016-11-14 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of Philips Spectral CT Applications supports viewing and analysis of images at energies selected from the available spectrum in order to provide information about... on . FDA classified it as Class II and gives the reason as "A software defect is causing issues with the IQon Spectral CT and Philips Spectral CT Applications while filming, saving and annotating images". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0706-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Spectral CT Applications supports viewing and analysis of images at energies selected from the available spectrum in order to provide information about the chemical composition of the body materials and/or contrast agents.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
A software defect is causing issues with the IQon Spectral CT and Philips Spectral CT Applications while filming, saving and annotating images
Root cause tag
software
FDA root cause
Software design
Action
A Field Safety Notice (Attachment #1 Field Safety Notice FSN 16-061-72800659 and FSN 88100051), will be sent to US consignees via certified mail. Philips Key Market representatives will distribute the notice internationally, based on local regulations. Additionally, a Field Service Engineer will contact each site to schedule a time to implement the Field Change Order.
Quantity in commerce
20
Distribution
Software was distributed in the following states: AR, IN, LA, MA, MN, OR and TN. Softward was also shipped to the following countries: Belgium, Denmark, France, Germany, Japan, Korea, Switzerland and United Kingdom.
Product codes and lots
50001 50002 50004 50005 50006 50008 50009 50100 50101 50102 50103 430000 430001 430002 430003 430004 430006 430007 430501 430502

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0706-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0705-2017Class IITerminatedVoluntary: Firm initiated
Z-0706-2017Class IITerminatedVoluntary: Firm initiated