510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0733-2017Class II

Cook Spectrum Minocycline/Rifampin Impregnated Double Lumen Central Venous Catheter Tray Cook Spectrum...

Cook Incorporated / initiated 2016-11-28 / Terminated

Cook Inc. began a recall of Cook Spectrum Minocycline/Rifampin Impregnated Double Lumen Central Venous Catheter Tray Cook Spectrum Minocycline/Rifampin Impregnated Triple Lumen Central... on . FDA classified it as Class II and gives the reason as "Cook Medical has received reports of the trays containing the incorrect needle, which in some cases could result in the inability to pass a wire guide through the needle. Potential adverse events that may occur as a...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0733-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Cook Spectrum Minocycline/Rifampin Impregnated Double Lumen Central Venous Catheter Tray Cook Spectrum Minocycline/Rifampin Impregnated Triple Lumen Central Venous Catheter Tray
Product code
FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Reason for recall
Cook Medical has received reports of the trays containing the incorrect needle, which in some cases could result in the inability to pass a wire guide through the needle. Potential adverse events that may occur as a result of utilizing needles of incorrect size include delay in treatment, which could be critical in an emergent situation. There have been no reports of illness or injury associated with this issue.
FDA root cause
Mixed-up of materials/components
Action
Cook Medical sent an Urgent - Medical Device Recall letter dated November 23, 2016. to all affected customers. The letter identified the product the problem and the action needed to be taken. Action to be taken: 1. Please review the attached list of affected products and lot numbers that were shipped to your account and quarantine any affected products that remain unused. 2. Immediately collect and return all unused affected products to Cook Medical as soon as possible for credit. 3. Please complete the attached Acknowledgement and Receipt Form and return it to Cook Medical. 4. Please report any adverse event to Cook Medical Customer Relations at (800) 457-4500 or 1(812) 339-2235, Monday through Friday between 7:30 am and 5:00 pm Eastern Standard Time or e-mail at [email protected]. Transmission of this notice: This notice must be passed on to appropriate personnel, including down to the user level, within your organization or to any organization where the potentially affected devices have been transferred. Should you have any questions or concerns, please do not hesitate to contact Cook Medical Customer Relations at (800) 457-4500 or 1(812) 339-2235.
Quantity in commerce
99 units
Distribution
AZ CA DC GA IL LA MN MS NC NV NY OK PA TN TX WI
Product codes and lots
catalog number: C-UDLMY-401J-ABRM-HC-FST Lot 6498570 catalog number: C-UTLMY-501J-ABRM-HC-FST Lot 6501835  Exp date 9/30/2018

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0733-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0733-2017Class IITerminatedVoluntary: Firm initiated