MEDTRONIC ARRIVE BRAIDED TRANSSEPTAL SHEATH 10F C90 S79CM D84CM GLOBAL Catalog # 990079-090-000 Is designed...
Oscor, Inc. began a recall of MEDTRONIC ARRIVE BRAIDED TRANSSEPTAL SHEATH 10F C90 S79CM D84CM GLOBAL Catalog # 990079-090-000 Is designed to facilitate the introduction of catheters to any... on . FDA classified it as Class II and gives the reason as "Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0817-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Oscor Inc.
- Product
- MEDTRONIC ARRIVE BRAIDED TRANSSEPTAL SHEATH 10F C90 S79CM D84CM GLOBAL Catalog # 990079-090-000 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Oscor sent two(2) Product Recall notifications: -HOSPITALS: Immediately check your inventory to confirm that you do or do not have units from these lot(s) in your possession. Identify and set aside any unit from the identified lot in a manner that ensures the affected product will not be used. Check all storage and usage. Review, complete, sign and return the enclosed Acknowledgement Form attached in the letter, directly to Oscor Inc. at the fax number or e-mail on the form, for immediate replacement (as applicable). Who to call: Please call Oscor Inc. Customer Service Department at 727-937-2511 or email [email protected] Hours of operations: Monday to Friday from 8:30AM to 5:30PM Eastern Time. - DISTRIBUTORS: Immediately check your inventory to confirm that you do or do not have units from these lot(s) in your possession. Identify and set aside any unit from the identified lot in a manner that ensures the affected product will not be used. Check all storage and usage. Please pull a list of the hospitals (end user) impacted by the affected lots and retrieve inventory for replacement, if available. Review, complete, sign and return the enclosed Acknowledgement Form (Exhibit A) attached in this letter, directly to Oscor Inc. at the fax number or e-mail on the form, for immediate replacement (as applicable). Who to call: Please call Oscor Inc. Customer Service Department at 727-937-2511 or email [email protected] Hours of operation: Monday to Friday from 8:30AM to 5:30PM Eastern Standard Time.
- Quantity in commerce
- 991
- Distribution
- Worldwide Distribution - US including CA, NJ, MD, TX, FL, NC, MO, and Puerto Rico. Internationally to AUSTRIA, BOLIVIA, CANADA, CHILE, CYPRUS, NETHERLANS, EGYPT, GERMANY, HONG KONG, ISRAEL, ITALY, SWITZERLAND, and VIETNAM
- Product codes and lots
- Lot #'s: C1-08341 Exp. November 11, 2016, C1-08607 Exp. February 20, 2017, C1-08803 Exp. April 3, 2017, C1-08941 Exp. May 19, 2017, C1-09173 Exp. August 25, 2017, C1-09394 Exp. November 16, 2017, C1-09531 Exp. December 29, 2017, C1-09718 Exp. February 11, 2018, C1-09748 Exp. February 16, 2018, C1-09899 Exp. March 12, 2018, C1-10040 Exp. April 5, 2018, C1-10149 Exp. April 23, 2018, C1-10330 Exp. May 11, 2018, C1-10329 Exp. May 11, 2018, C1-10382 Exp. May 17, 2018, C1-11571 Exp. September 25, 2018, and OR-04757 Exp. December 22, 2018 53
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0790-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0791-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0792-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0793-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0794-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0795-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0796-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0797-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0798-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0799-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0800-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0801-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0802-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0803-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0804-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0805-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0806-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0807-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0808-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0809-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0810-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0811-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0812-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0813-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0814-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0815-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0816-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0817-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0818-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0819-2017 | Class II | Terminated | Voluntary: Firm initiated |