Z-0918-2017Class II

Shimadzu X-ray High Voltage Generator Model No. UD150B-40 on SONIALVISION safire II/17 systems with...

Shimadzu Medical Systems / initiated 2016-11-20 / Terminated

Shimadzu Medical Systems began a recall of Shimadzu X-ray High Voltage Generator Model No. UD150B-40 on SONIALVISION safire II/17 systems with 0.7/1.2JG326D X-Ray Tube Diagnostic x-ray high voltage... on . FDA classified it as Class II and gives the reason as "Shimadzu Medical Systems is recalling the Shimadzu X-ray High Voltage Generator due to the possibility of oil leaking from the x-ray tube housing under certain conditions". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0918-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Shimadzu Medical Systems
Product
Shimadzu X-ray High Voltage Generator Model No. UD150B-40 on SONIALVISION safire II/17 systems with 0.7/1.2JG326D X-Ray Tube Diagnostic x-ray high voltage generator.
Product code
IZO Generator, High-Voltage, X-Ray, Diagnostic
Reason for recall
Shimadzu Medical Systems is recalling the Shimadzu X-ray High Voltage Generator due to the possibility of oil leaking from the x-ray tube housing under certain conditions.
FDA root cause
Under Investigation by firm
Action
The firm, Shimadzu Medical Systems USA, sent an "Urgent Medical Device Recall notice" and Field Safety Notice dated 11/20/16 to customers to inform them that Shimadzu Corporation has identified a potential issue in the Shimadzu X-ray High Voltage Generator UD150B-40. Customers are informed of how to identify the affected products, corrective action plan by Shimadzu, and the actions to be taken by customers. Customers that have not received a call or visit from Shimadzu are instructed to call Shimadzu Medical Systems National Technical Support at (800) 228-1429. Shimadzu sent an updated "URGENT Voluntary Medical Device Recall Notice" letter dated 11/30/16 to its customers on 11/30/16 to inform them that Shimadzu Corp. has identified a potential issue in the Shimadzu X-ray High Voltage Generator UD150B-40 on SONIALVISION safire II/17 systems with 0.7/1.2JG326D X-Ray Tube. The letter informs the customers that when operated for prolonged periods of time at low rate serial exposure, the heat load capacity of the X-ray anode disc (target) may become damaged and leak insulating oil. Customers with questions are instructed to contact Shimadzu National Technical Support at (800) 228-1429.
Quantity in commerce
13 units
Distribution
US Distribution to states of: AZ, WA, CA, LA, FL, SC, and NJ.
Product codes and lots
Serial No. 3M5249D1C013 3M5249D22013 3M5249D29002  3M5249D2A004 3M5249D2C004   3M5249D31009 3M5249D32012 3M5249D33003  3M5249D34001 3M5249D35002 3M5249D37006 3M5249D41006 3M5249D46003

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0918-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0918-2017Class IITerminatedVoluntary: Firm initiated