Z-0731-2017Class II
MOSAIQ Oncology Information System The product is used to manage workflows for treatment planning and...
Elekta Limited / initiated 2016-12-02 / Terminated / root cause: software
Elekta, Inc. began a recall of MOSAIQ Oncology Information System The product is used to manage workflows for treatment planning and delivery. It supports information flow among healthcare... on . FDA classified it as Class II and gives the reason as "Edits to Particle field definition parameters may not be saved when the field definition window is saved". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0731-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- MOSAIQ Oncology Information System The product is used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be use wherever radiotherapy and chemotherapy are prescribed.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Edits to Particle field definition parameters may not be saved when the field definition window is saved.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm, Elekta, sent an "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" letter dated 11/2016 was sent to affected customers on 12/2/2016. The letter describes the product, problem and actions to be taken. The customers were instructed to save and re-open the field definition form when manual edits have been performed to confirm that the edits were save; post this notice in a place accessible to all users; advise the appropriate personnel working with the product the content of the letter; and complete and return the Acknowledgement Form via fax to: 408 830 8003 immediately upon receipt but no later than 30 days. Elekta will resolve the problem with the MOSAIQ 2.60 SP7F, MOSAIQ 2.60 SP14 and 2.64 SP3. Customers are encouraged to contact Elekta support to get the corrected version. If you have any questions, contact RA Manager Americas at 1-770-300-9725 or email to: [email protected].
- Quantity in commerce
- 47
- Distribution
- Worldwide Distribution-US (nationwide) to states of: FL, IL, MO, NJ, OH, OK, TN, TX, VA, and countries of: Italy, Japan, South Korea, Sweden and United Kingdom.
- Product codes and lots
- Versions 2.60 an higher
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0731-2017 | Class II | Terminated | Voluntary: Firm initiated |