Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0881-2017Class II

K-Wire 1.4mm x 12in Non-Threaded, Part Number 9080-12U; K-Wire Threaded 1.4mm x 18, Part Number 9080-18T...

Spinal Elements, Inc / initiated 2016-11-22 / Terminated / root cause: manufacturing process

Amendia, Inc began a recall of K-Wire 1.4mm x 12in Non-Threaded, Part Number 9080-12U; K-Wire Threaded 1.4mm x 18, Part Number 9080-18T; K-Wire Non-Threaded 1.4mm x 18, Part Number 9080-18U... on . FDA classified it as Class II and gives the reason as "The rods are packaged in packages with unacceptable seals. In some cases, the integrity of the seal is compromised by channels or voids". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0881-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Spinal Elements, Inc
Product
K-Wire 1.4mm x 12in Non-Threaded, Part Number 9080-12U; K-Wire Threaded 1.4mm x 18, Part Number 9080-18T; K-Wire Non-Threaded 1.4mm x 18, Part Number 9080-18U; 1.4mm x 24 Threaded Guide Wire, Part Number 9080-24T; K-Wire Non-Threaded 24, Part Number 9080-24U; Blunt K-Wire, 18 inches, Threaded, Part Number 9080B-18T; Blunt K-Wire, 18 inches, Unthreaded, Part Number 9080B-18U; Blunt K-Wire, 24 inches, Unthreaded, Part Number 9080B-24U; 1.4mm x 18 Blunt Threaded Nitinol Guide Wire, Part Number 9080B-N-18T; 1.4mm x 18" Blunt Non-Threaded Nitinol Guide Wire, Part Number 9080B-N-18U; Lodestar K-wire Non-Threaded, 18, Part Number 9080L-18U; 1.4mm x 18 Threaded Nitinol Guide Wire, Part Number 9080-N-18T; 1.4mm x 18 Non-Threaded Nitinol Guide Wire, Part Number 9080-N-18U The guide wire is passed through the cannulated handle of the Dual Stylet Needles and pressed into the bone
Product code
HTY Pin, Fixation, Smooth
Reason for recall
The rods are packaged in packages with unacceptable seals. In some cases, the integrity of the seal is compromised by channels or voids.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Amendia sent an Urgent Medical Device Recall letter dated November 22, 2016, to all affected customers. The letter requested that customers examine their inventory and quarantine the product. The letter also requested that they contact Amendia to confirm receipt of the letter and to obtain a Return Material Authorization for any product on hand. Customers with questions were instructed to call the following numbers: 770-575-5224 770-575-5236 770-575-5369 For questions regarding this recall call 404-610-7215.
Quantity in commerce
5833 units
Distribution
Nationwide Distribution to AL, AZ, AR, CA, CO, FL, GA, IA, KS, KY, LA, MA, MI, MN, MO, NV, NJ, NC, SC, TN, TX, VA, WA, WV, WI and Puerto Rico
Product codes and lots
Part Number 9080-12U Lot Number 121306, 121307, 122628, 122721, 124400, 132732, 135403; Part Number 9080-18T Lot Number 122486, 122626, 123957, 124408, 127435, 129132, 129803, 129935, 131072, 135560, 135566, 135567; Part Number 9080-18U Lot Number 122470, 123948, 129802, 129936, 130390, 130391, 131073; Part Number 9080-24T Lot Number 133359; Part Number 9080-24U Lot Number 134108, 134112, 134908; Part Number 9080B-18T Lot Number 124547, 126820, 127437, 131454, 132592, 134181, 135401; Part Number 9080B-18U Lot Number 122468, 123938, 126504, 127436, 129800, 130947, 131453, 132750, 133645, 134169, 134907, 135563, 135578, 135579, 135580; Part Number 9080B-24U Lot Number 122239, 130392, 130393, 131839, 133988, 135359, 135360, 135400; Part Number 9080B-N-18T Lot Number 124137, 125331, 128472, 130612, 132229, 135584, 135585, 137464; Part Number 9080B-N-18U Lot Number 123556, 125199, 126821, 128348, 128469, 129443, 129915, 130836, 131626, 133118, 134180, 135208; Part Number 9080L-18U Lot Number 127877, 128351, 130389, 130946, 133261, 135210, 135593; Part Number 9080-N-18T Lot Number 122488, 123863, 130914; Part Number 9080-N-18U Lot Number 122487, 123564, 124402, 126125, 129005, 129077, 133499, 135562, 137462.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0881-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0881-2017Class IITerminatedVoluntary: Firm initiated
Z-0882-2017Class IITerminatedVoluntary: Firm initiated
Z-0883-2017Class IITerminatedVoluntary: Firm initiated