Z-0919-2017Class II

SOMATOM Scope, Computed tomography x-ray system

Siemens Medical Solutions USA, Inc. / initiated 2016-11-22 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of SOMATOM Scope, Computed tomography x-ray system on . FDA classified it as Class II and gives the reason as "Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0919-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
SOMATOM Scope, Computed tomography x-ray system
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens mailed on December 01, 2016 a Customer Safety Notice to affected customers to inform them of a corrective action that is being performed to prevent a possible hazard to persons and equipment. The notice explained the underlying issue, the effect the issue has on the system, the associated risks and actions they will take to correct the issue. In addition, customers will receive a Customer Safety Notice via Hand Delivery from the Siemens Customer Service Engineer at the scheduled appointment for the corrective installation.
Quantity in commerce
13 systems
Distribution
Distributed to: IL IL GA SC SD OK NY MI IN MD AR
Product codes and lots
Model # 10967666 Serial Number:59376 91766 91767 115456 115454 59380 77560 98840 77557 91758 59377 59378 115451

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0919-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0919-2017Class IITerminatedVoluntary: Firm initiated
Z-0920-2017Class IITerminatedVoluntary: Firm initiated
Z-0921-2017Class IITerminatedVoluntary: Firm initiated