SOMATOM Perspective, Computed tomography x-ray system
Siemens Medical Solutions USA, Inc began a recall of SOMATOM Perspective, Computed tomography x-ray system on . FDA classified it as Class II and gives the reason as "Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0921-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- SOMATOM Perspective, Computed tomography x-ray system
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Siemens is providing a new installation of the Residual Current Monitor (RCM) in the Line Connection Box (LCB) to resolve a potential defect that could lead to a power cutout during use of the SOMATOM Scope, Scope Power and Perspective systems.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Siemens mailed on December 01, 2016 a Customer Safety Notice to affected customers to inform them of a corrective action that is being performed to prevent a possible hazard to persons and equipment. The notice explained the underlying issue, the effect the issue has on the system, the associated risks and actions they will take to correct the issue. In addition, customers will receive a Customer Safety Notice via Hand Delivery from the Siemens Customer Service Engineer at the scheduled appointment for the corrective installation.
- Quantity in commerce
- 13 systems
- Distribution
- Distributed to: IL IL GA SC SD OK NY MI IN MD AR
- Product codes and lots
- Model # 10967888 Serial Number:59376 91766 91767 115456 115454 59380 77560 98840 77557 91758 59377 59378 115451
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0919-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0920-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0921-2017 | Class II | Terminated | Voluntary: Firm initiated |