Z-0939-2017Class II
DR Systems Unity PACS software, now known as Merge Unity PACS software
Merge Healthcare Incorporated / initiated 2015-10-15 / Terminated / root cause: software
Merge Healthcare, Inc. began a recall of DR Systems Unity PACS software, now known as Merge Unity PACS software on . FDA classified it as Class II and gives the reason as "The software fails to associate to the correct MG image if there are two images for the same view". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0939-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merge Healthcare Incorporated
- Product
- DR Systems Unity PACS software, now known as Merge Unity PACS software.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- The software fails to associate to the correct MG image if there are two images for the same view.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Merge sent an Urgent : Medical Device Recall letter dated July 11, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Be aware of the following workaround: Customer I User: Ensure both" tome" images are reviewed when more than one is captured for the same view. Merge has a released fix available for this issue. YOUR RESPONSE TO THIS NOTIFICATION IS REQUIRED Please reply using the enclosed form and the return addressed envelope. Your response is required no later than August 2,2016. For further questions regarding this recall please call (877) 741-5369
- Quantity in commerce
- 9 sites potentially have the effected software
- Distribution
- Distribution was made to medical facilities located in MT, CA, PA, and TX.
- Product codes and lots
- Version 11.0
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0939-2017 | Class II | Terminated | Voluntary: Firm initiated |