Z-1076-2017Class II

Mallinckrodt Pharmaceuticals INOmax DSIR Plus delivery system. Model 10007. Delivers INOMAX (nitric oxide for...

Mallinckrodt Manufacturing LLC / initiated 2016-11-21 / Terminated / root cause: manufacturing process

INO Therapeutics (dba Ikaria) began a recall of Mallinckrodt Pharmaceuticals INOmax DSIR Plus delivery system. Model 10007. Delivers INOMAX (nitric oxide for inhalation). The INOmax DSIR uses a dual-channel... on . FDA classified it as Class II and gives the reason as "Mallinckrodt discovered that some devices may have been manufactured with an out-of-tolerance manometer which can result in in an inaccurate proportional valve calibration table in an affected device". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1076-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Mallinckrodt Manufacturing LLC
Product
Mallinckrodt Pharmaceuticals INOmax DSIR Plus delivery system. Model 10007. Delivers INOMAX (nitric oxide for inhalation). The INOmax DSIR uses a dual-channel design to ensure the safe delivery of INOMAX. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and 02 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOMAX delivery independent of monitoring but also allows the monitoring system to shutdown INOMAX delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm. The delivery system can also shut down delivery if it detects certain serious problems with the monitoring system.
Product code
MRN Apparatus, Nitric Oxide Delivery
Reason for recall
Mallinckrodt discovered that some devices may have been manufactured with an out-of-tolerance manometer which can result in in an inaccurate proportional valve calibration table in an affected device.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
Recall letter was sent to customers in a e-mail beginning November 21, 2016 with a had copy sent via FedEx overnight with signature receipt beginning December 5, 2016. Devices will be picked up and replaced by Mallinckrodt. Questions can be directed to 1-877-KNOW-INO (1-877-566-9466).
Quantity in commerce
45
Distribution
US Distribution to the states of : CA, CO, DC, FL, GA, HI, IA, IL, IN, MI, MN, NC, NJ, NM, NY, OH, PA, and TX
Product codes and lots
Serial Numbers: DS20070030, DS20070043, DS20070094, DS20070268, DS20090096, DS20090458, DS20090484, DS20090539, DS20090620, DS20090645, DS20100071, DS20100118, DS20100216, DS20100266, DS20100268, DS20100330, DS20100439, DS20100496, DS20100604, DS20100675, DS20100744, DS20100762, DS20100950, DS20101199, DS20101337, DS20101359, DS20101369, DS20101406, DS20101443, DS20101524, DS20110242, DS20110480, DS20110577, DS20110849, DS20110868, DS20110912, DS20110915, DS20110972, DS20110976, DS20110990, DS20120008, DS20120161, DS20120266, DS20120281, DS20120370.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1076-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1076-2017Class IITerminatedVoluntary: Firm initiated