Z-0945-2017Class II

Modular SMF(TM) The Modular SMF(TM) and Modular REDAPT(TM) Hip Systems are indicated for individuals...

Smith & Nephew, Inc. / initiated 2016-11-15 / Terminated / root cause: design

Smith & Nephew, Inc. began a recall of Modular SMF(TM) The Modular SMF(TM) and Modular REDAPT(TM) Hip Systems are indicated for individuals undergoing primary and revision surgery where other... on . FDA classified it as Class II and gives the reason as "All lots of modular neck hip prostheses are being recalled due to a higher than anticipated complaint and adverse event trend". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0945-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
Modular SMF(TM) The Modular SMF(TM) and Modular REDAPT(TM) Hip Systems are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma or non-inflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant. Hip components are also indicated for inflammatory degenerative joint disease including rheumatoid arthritis, arthritis secondary to a variety of diseases and anomalies, and congenital dysplasia; treatments of nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; endoprosthesis, femoral osteotomy, or Girdlestone resection; fracture-dislocation of the hip; and correction of deformity. The Modular SMF(TM) and Modular REDAPT(TM) Hip Systems are intended for single use only and are to be implanted without bone cement.
Product code
MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
Reason for recall
All lots of modular neck hip prostheses are being recalled due to a higher than anticipated complaint and adverse event trend.
Root cause tag
design
FDA root cause
Device Design
Action
Smith & Nephew sent an Urgent Product Recall Notification letter dated November 15, 2016, to all affected customers. Customers were instructed to inspect their inventory and locate any unused devices and quarantine immediately. Complete the Inventory Return Certification Form and obtain a return authorization (RA) number if there was product to return. For questions regarding this recall call 978-749-1440.
Quantity in commerce
11999 units
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Austria, Australia, Belgium, Canada, Switzerland, China, Cyprus, Czech Republic, Dubai, Germany, Denmark, Spain, France, United Kingdom, Greece, Hong Kong, India, Italy, Mexico, Netherlands, Poland, Puerto Rico, Portugal, Russia, Turkey, and South Africa.
Product codes and lots
71352401 SMF STEM RSA SIZE 1 71352402 SMF STEM RSA SIZE 2 71352403 SMF STEM RSA SIZE 3 71352404 SMF STEM RSA SIZE 4 71352405 SMF STEM RSA SIZE 5 71352406 SMF STEM RSA SIZE 6 71352407 SMF STEM RSA SIZE 7 71352408 SMF STEM RSA SIZE 8 71352409 SMF STEM RSA SIZE 9 71352501 SMF STEM WITH STIKTITE SZ 1 71352502 SMF STEM WITH STIKTITE SZ 2 71352503 SMF STEM WITH STIKTITE SZ 3 71352504 SMF STEM WITH STIKTITE SZ 4 71352505 SMF STEM WITH STIKTITE SZ 5 71352506 SMF STEM WITH STIKTITE SZ 6 71352507 SMF STEM WITH STIKTITE SZ 7 71352508 SMF STEM WITH STIKTITE SZ 8 71352509 SMF STEM WITH STIKTITE SZ 9

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0945-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0944-2017Class IITerminatedVoluntary: Firm initiated
Z-0945-2017Class IITerminatedVoluntary: Firm initiated
Z-0946-2017Class IITerminatedVoluntary: Firm initiated