510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1020-2017Class II

Reunion TSA Peg Alignment Sound Catalog #5901-1038 The Peg Alignment Sound (PIN 5901-1038, Revision A) is an...

Howmedica Osteonics Corp / initiated 2016-12-02 / Terminated / root cause: design

Stryker Howmedica Osteonics Corp. began a recall of Reunion TSA Peg Alignment Sound Catalog #5901-1038 The Peg Alignment Sound (PIN 5901-1038, Revision A) is an instrument used in the ReUnion TSA shoulder system... on . FDA classified it as Class II and gives the reason as "Reunion TSA Peg Alignment Sound broke during surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1020-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Reunion TSA Peg Alignment Sound Catalog #5901-1038 The Peg Alignment Sound (PIN 5901-1038, Revision A) is an instrument used in the ReUnion TSA shoulder system for confirming seating height and peg alignment of the TSA Self-Pressurizing Polyethylene Glenoid implant prior to final implantation into the glenoid vault. Its purpose is to provide assurance that the drilled holes for seating of the Glenoid implant have been properly prepared using the provided instrumentation
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
Reunion TSA Peg Alignment Sound broke during surgery.
Root cause tag
design
FDA root cause
Device Design
Action
Stryker Orthopaedics sent an Urgent Medical Device Recall Notification letter dated December 2, 2016, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to inform users of the Urgent Medical Device Recall and forward the notice to all those individuals who need to be aware within their organization. Return all affected products available to: Stryker Orthopaedics/PFA Product Returns Attn: Distribution Inventory Team 325 Corporate Drive Dock M-East Mahwah, NJ 07431 1360184/RA2016-112 Complete and sign the enclosed Business Reply Form and fax a copy to 855-261-3635 or email to [email protected]. Customers with questions were instructed to call 201-831-6693.
Quantity in commerce
131 units
Distribution
Nationwide Distribution
Product codes and lots
Catalog #5901-1038

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1020-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1020-2017Class IITerminatedVoluntary: Firm initiated