Z-0878-2017Class II

Merge OrthoCase software. The firm name on the label is Merge Healthcare, Hartland, WI

Merge Healthcare Incorporated / initiated 2015-10-08 / Terminated / root cause: software

Merge Healthcare, Inc. began a recall of Merge OrthoCase software. The firm name on the label is Merge Healthcare, Hartland, WI on . FDA classified it as Class II and gives the reason as "Measurements in the software are changing after saving a plan with a measurement result, which may result in potential patient injury or delay in diagnosis or treatment". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0878-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merge Healthcare Incorporated
Product
Merge OrthoCase software. The firm name on the label is Merge Healthcare, Hartland, WI.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Measurements in the software are changing after saving a plan with a measurement result, which may result in potential patient injury or delay in diagnosis or treatment.
Root cause tag
software
FDA root cause
Software design
Action
Merge issued a letter dated March 28, 2016, via email on April 4, 2016, or via certified mail if they did not have an email address for the customer. The letter notifies the customer of the issue, informs them of a workaround and that a fix is not yet available but they will be notified once one is available. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. The notification also informs the customer their response is required using the enclosed form and the return addressed envelope. The response form asks, in part, if they are interested in accepting the fix. If they decline the fix, they are asked to provide the reason why. The response was required no later than April 18, 2016. Nonresponders were issued a second letter dated July 11, 2016, asking for a response by August 2, 2016. For further questions, please call (877) 741-5369.
Quantity in commerce
30 sites potentially have the effected versions
Distribution
Distribution was made to medical facilities in CA, CO, CT, MA, MD, MN, MT, NE, NV, NY NJ, PA, SC, WA, and WI. Foreign distribution was made to Canada, as well as other countries.
Product codes and lots
Versions: 2.0, 2.1, 2.1.1, 3.1, 3.2, 3.2.4, 3.3, 3.3.1, 3.4, 3.5.1, 3.5.1.1, 3.5.2, 3.5.3, 3.5.4, 3.5.5, 3.6, 3.6.1, 3.6.2, 3.7, 3.7.1, 3.7.2, 3.7.2 Vet, and 3.7.3

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0878-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0878-2017Class IITerminatedVoluntary: Firm initiated