Merge Eye Station Import Utility (ESIU) when used with Merge Eye Station and Merge Eye Care PACS. The firm...
Merge Healthcare, Inc. began a recall of Merge Eye Station Import Utility (ESIU) when used with Merge Eye Station and Merge Eye Care PACS. The firm name on the labeling is Merge Healthcare, Hartland... on . FDA classified it as Class II and gives the reason as "Eye Station images were not importing properly and were imported under "unknown" due to an issue when validating patients using only an Medical Record Number (MRN)". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1246-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merge Healthcare Incorporated
- Product
- Merge Eye Station Import Utility (ESIU) when used with Merge Eye Station and Merge Eye Care PACS. The firm name on the labeling is Merge Healthcare, Hartland, WI.
- Product code
- NFJ System, Image Management, Ophthalmic
- Reason for recall
- Eye Station images were not importing properly and were imported under "unknown" due to an issue when validating patients using only an Medical Record Number (MRN)
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The recalling firm issued a letter dated 3/28/2016 via e-mail. If no email address was available, the letters were issued via certified mail. A follow-up letter dated 7/11/2016 was also issued.
- Quantity in commerce
- 1,627 sites potentially have the affected versions
- Distribution
- Distribution was nationwide to medical facilities. There was also government and foreign distribution, including Canada.
- Product codes and lots
- Merge ESIU: Versions 1.0, 1.1.1, 1.1.2, and 1.1.3; Merge Eye Station: All versions; Merge Eye Care PACS: Versions 4.1, 4.2, 4.3, 5.0, and 5.1
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1246-2017 | Class II | Terminated | Voluntary: Firm initiated |