510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0923-2017Class II

Allura Xper FD10 722026 is intended for: Cardiovascular and vascular X-ray imaging applications, including...

Philips North America LLC / initiated 2016-12-01 / Terminated / root cause: manufacturing process

Philips Electronics North America Corporation began a recall of Allura Xper FD10 722026 is intended for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA... on . FDA classified it as Class II and gives the reason as "Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0923-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Allura Xper FD10 722026 is intended for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and Electro Physiology(EP).
Product code
IZI System, X-Ray, Angiographic
Reason for recall
Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Customers were notified of the recall via letter sent on 12/1/16. The letter explains the issue and the action that Philips plans to take.
Quantity in commerce
1
Distribution
US: WA OUS: China, Spain, Korea, Republic of, Germany
Product codes and lots
783

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0923-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0922-2017Class IITerminatedVoluntary: Firm initiated
Z-0923-2017Class IITerminatedVoluntary: Firm initiated
Z-0924-2017Class IITerminatedVoluntary: Firm initiated
Z-0925-2017Class IITerminatedVoluntary: Firm initiated