Z-1005-2017Class II

FV-RF (Clinix RF Plus)

Philips North America LLC / initiated 2016-11-18 / Terminated / root cause: design

Philips Medical Systems (Cleveland) Inc began a recall of FV-RF (Clinix RF Plus) Product Usage: A Stationary X-ray System, a device intended for radiographic/fluoroscopic examinations of various anatomical regions... on . FDA classified it as Class II and gives the reason as "There's a potential for liquid penetration into the back side of the front control panel of the X-ray diagnostic table, which can cause a short circuit of the terminals of the tilt operation and cause an unintentional...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1005-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
FV-RF (Clinix RF Plus) Product Usage: A Stationary X-ray System, a device intended for radiographic/fluoroscopic examinations of various anatomical regions. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
Product code
KPR System, X-Ray, Stationary
Reason for recall
There's a potential for liquid penetration into the back side of the front control panel of the X-ray diagnostic table, which can cause a short circuit of the terminals of the tilt operation and cause an unintentional tilt of the table.
Root cause tag
design
FDA root cause
Component design/selection
Action
Philips sent an Urgent Field Safety Notice Medical Device Correction letter dated December 2, 2016 to customers via USPS certified letter with returned receipt. The letter identified the affected product, problem and actions to be taken. Additionally, your local Philips representative has been provided details regarding this Device Correction and can be contacted if you need any further information or support concerning this issue: Technical Support Line: 1-800-722-9377
Quantity in commerce
41
Distribution
US Nationwide distribution in the following states: AL, CA, FL, GA, HI, IL, IN, KS, LA, MA, MI, MS, NJ, NY, OH, PA, SC, VA, WI, WV, & WY.
Product codes and lots
System Serial Numbers:  0162R15403 0162A66105 0162A65401 218134 217530 209677 716334 217721 0162R16001 0162A65301 0162R15803 0162R16002 209669 0162R15504 0162R15602 0162A66104 0162R16007 0162R15802 0162R15702 0162R16003 212866 218047 0162A66005 0162R15603 0162R16004 218085 0162R15804 217296 0162A65402 0162P21004

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1005-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1005-2017Class IITerminatedVoluntary: Firm initiated