PreludeEASE Hydrophilic Sheath Introducer Dilator The Merit PreludeEASE Hydrophilic Sheath Introducer is...
Merit Medical Systems, Inc. began a recall of PreludeEASE Hydrophilic Sheath Introducer Dilator The Merit PreludeEASE Hydrophilic Sheath Introducer is intended to provide access and facilitate the... on . FDA classified it as Class II and gives the reason as "Recalling dilators included with the 6F PreludeEASE Hydrophilic Sheath Introducer because the dilator tips were manufactured with an undersized inner diameter and therefore will not allow passage of an 0.021 (0.53 mm)...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0949-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merit Medical Systems, Inc.
- Product
- PreludeEASE Hydrophilic Sheath Introducer Dilator The Merit PreludeEASE Hydrophilic Sheath Introducer is intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries, including but not limited to the radial artery, while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. The access needle with inner metal needle and outer plastic cannula is used to gain access to the vein or artery for placement of guide wires
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Recalling dilators included with the 6F PreludeEASE Hydrophilic Sheath Introducer because the dilator tips were manufactured with an undersized inner diameter and therefore will not allow passage of an 0.021 (0.53 mm) guide wire.
- Root cause tag
- manufacturing process
- FDA root cause
- Process change control
- Action
- Merit Medical sent an Urgent Product Recall Notice dated December 1, 2016, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to identify any affected devices within their facility, quarantine them, and discontinue use. Ensure that all personnel to whom the devices were distributed are made aware of this field action. Consignees were asked to work with their Merit Sales Representative to arrange product return. Consignees with questions were instructed to contact their Merit Sales Representative or Merit Customer Service at [email protected] or at 801-208-4381. For questions regarding this recall call 801-208-4623.
- Quantity in commerce
- 335
- Distribution
- Nationwide Distribution
- Product codes and lots
- Lots: H1014678, H1039272, H1029150, H1039269, H1029151
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0949-2017 | Class II | Terminated | Voluntary: Firm initiated |