Z-0910-2017Class II
Precision 500D
GE Healthcare / initiated 2016-11-21 / Terminated / root cause: manufacturing process
GE Healthcare, LLC began a recall of Precision 500D on . FDA classified it as Class II and gives the reason as "GE Healthcare is conducting a recall due to the potential fall of aged CRT monitors - Prestige 2, Prestilix, Prestige VH, Prestige SI, RFX/SFX, Legacy, Precision 500D, ADVANTX and Innova 2000 systems". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0910-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Precision 500D
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- GE Healthcare is conducting a recall due to the potential fall of aged CRT monitors - Prestige 2, Prestilix, Prestige VH, Prestige SI, RFX/SFX, Legacy, Precision 500D, ADVANTX and Innova 2000 systems.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Consignees were sent on 11/21/2016 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# FMI 10913 dated November 21, 2016. The letter was addressed to Director of Radiology, Risk Manager/Hospital Administrator, and Director of Biomedical Engineering. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Quantity in commerce
- Total for all products 1,598 (USA 824, OUS 774)
- Distribution
- US: AL, AR, AZ, CA, CO, CT, DC, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT,NC, NE NH, NJ, NM, NY, NV, OH, OK, PA, RI, SC, SD, TN, TX,, VA, WA, WO, WY, WV. OUS: Algeria Armenia Australia Austria Brazil Canada Chile China Denmark Egypt France Germany Guatemala Honduras Hong Kong Hungary Iceland India Ireland Italy Japan Korea Libya Mexico Netherlands Netherlands Peru Poland Portugal Russia Saudi Arabia Serbia Sierra Leone Slovakia Spain Taiwan Thailand Turkey United Arab Emirates United Kingdom Venezuela Vietnam
- Product codes and lots
- Mfg. Lot or Serial # System ID 00000034206YY3 229353TR2 00000146304MO1 423778ERM4 Not Available 817255NHPD1
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0904-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0905-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0906-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0907-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0908-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0909-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0910-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0911-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0912-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0913-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0914-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0915-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0916-2017 | Class II | Terminated | Voluntary: Firm initiated |