MiniLoc Safety Infusion Set Winged without Y-Injection Site
Bard Access Systems began a recall of MiniLoc Safety Infusion Set Winged without Y-Injection Site. Product Code: S02022-75. Packaged as single units in thermoformed trays. There are 20 MiniLoc... on . FDA classified it as Class II and gives the reason as "Bard Access Systems announces a voluntary field action for the SafeStep Huber Needle Set with Y-Injection Site and MiniLoc Safety Infusion Set Winged without Y-Injection Site devices because sets may contain one or more...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0985-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bard Access Systems, Inc.
- Product
- MiniLoc Safety Infusion Set Winged without Y-Injection Site. Product Code: S02022-75. Packaged as single units in thermoformed trays. There are 20 MiniLoc trays per case and 160 per carton. Product Usage: Intended for the administration of fluids and drugs, or blood sampling through surgically implanted vascular ports.
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- Bard Access Systems announces a voluntary field action for the SafeStep Huber Needle Set with Y-Injection Site and MiniLoc Safety Infusion Set Winged without Y-Injection Site devices because sets may contain one or more units where the needle length and gage differs from the length and gage described on the label.
- Root cause tag
- labeling and UDI
- FDA root cause
- Mixed-up of materials/components
- Action
- Bard sent an Urgent Notification letter dated December 16, 2016 to customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to identify and return affected product. For questions call 1-800-290-1689.
- Quantity in commerce
- 7,200 units
- Distribution
- Worldwide Distribution - US Nationwide in the states of CA, CO, FL, GA, HI, IL, MA, MD, NJ, OH, OR, TX.
- Product codes and lots
- Lot Numbers: ASAQS0343 & ASZIS027
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0985-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0986-2017 | Class II | Terminated | Voluntary: Firm initiated |