Z-0972-2017Class II

Artis Q/Q.zen, Angiographic x-ray system Stand alone system The Artis systems are a family of dedicated...

Siemens Medical Solutions USA, Inc. / initiated 2016-12-08 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Artis Q/Q.zen, Angiographic x-ray system Stand alone system The Artis systems are a family of dedicated angiography systems developed for single and biplane... on . FDA classified it as Class II and gives the reason as "Siemens initiated a corrective action to address two possible, mutually independent causes of a system defect related to the following: - In Artis Systems with A100 Plus or A100G generators, an attempt to resume...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0972-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Artis Q/Q.zen, Angiographic x-ray system Stand alone system The Artis systems are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee/ zeego and Q/ Q.zen include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee/ zeego and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Siemens initiated a corrective action to address two possible, mutually independent causes of a system defect related to the following: - In Artis Systems with A100 Plus or A100G generators, an attempt to resume operation following detection of a fault can result in the failure of a module in the high-voltage generator. - For biplane systems delivered with software version VD11B since April of this year, software problems in conjunction with graphics card may, in rare cases, result in the loss of image display in the examination room.
Root cause tag
software
FDA root cause
Software in the Use Environment
Action
Siemens mailed a Customer Safety Advisory Notice (CSAN) dated December 8, 2016, to all affected customers with Customer Safety Advisory Notice AX002/16/S. Additionally, Siemens has initiated a software update via Update Instruction AX001/16/S. Their Service Organization will contact each customer to schedule an appointment to perform the corrective action; however, customers can make contact for an earlier appointment. Customers with questions should call 1-888-826-9702. For questions regarding this recall call 610-448-6461.
Quantity in commerce
1,500 distributed Worldwide
Distribution
Nationwide Distribution
Product codes and lots
Model # 10848280, 10848281, 10848282, 10848353, 10848354, 10848355

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0972-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0971-2017Class IITerminatedVoluntary: Firm initiated
Z-0972-2017Class IITerminatedVoluntary: Firm initiated