Z-0990-2017Class II

Eclipse Treatment Planning System version 13.MR2 [13.06.31 with Smart Segmentation Knowledge Based Contouring...

Varian Medical Systems, Inc. / initiated 2016-12-02 / Terminated / root cause: design

Varian Medical Systems, Inc. began a recall of Eclipse Treatment Planning System version 13.MR2 [13.06.31 with Smart Segmentation Knowledge Based Contouring version 2.3 [2.3.12] Radiology: The Eclipse... on . FDA classified it as Class II and gives the reason as "Modifications in version 13.6MR2 for Contouring, SmartAdapt, and SmartSegmentation workspaces resulted in contours not being saved consistently in Eclipse. Treatment Planning System. The issue only occurs if certain...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0990-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Varian Medical Systems, Inc.
Product
Eclipse Treatment Planning System version 13.MR2 [13.06.31 with Smart Segmentation Knowledge Based Contouring version 2.3 [2.3.12] Radiology: The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Modifications in version 13.6MR2 for Contouring, SmartAdapt, and SmartSegmentation workspaces resulted in contours not being saved consistently in Eclipse. Treatment Planning System. The issue only occurs if certain conditions are fulfilled.
Root cause tag
design
FDA root cause
Device Design
Action
Varian Medical Systems sent an Urgent Medical Device Correction Letter dated December 5, 2016 , that were sent to all 5 affected sites on December 5, 2016 by a trackable method. Letters advised users of the issue and the specific scenario that must take place for the issue to occur. Varian is investigating the issue and will provide a correction once it is available. We sincerely apologize for any inconvenience and thank you in advance for your cooperation. If you require further clarification, please feel free to contact your local Varian Medical Systems Customer Support District or Regional Manager. This notice has been provided to the appropriate Regulatory Authorities. For further questions please call (650) 424-6833. Varian Oncology Help Desk Contact Information: Phone: +1.888.827.4265 Email: support @varian.com
Quantity in commerce
42 devices are affected
Distribution
US Distribution to the states of : NJ, TN. OR and FL.
Product codes and lots
Model # H48  Serial numbers: H48L574 H48L575 H48L576 H48L577 H48L578 H48L579 H48L580 H48L581 H48L582 H48L583 H48L584 H48L585 H48L586 H48L587 H482484 H487302 H48A587 H48D957 H480113 H480118 H481884 H481885 H481886 H481916 H481917 H484198 H484199 H488949 H48A680 H48A871 H48A872 H48A873 H48A874 H48A875 H48A876 H48A877 H48A878 H48A879 H48I531 H481195 H481236 H481237.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0990-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0990-2017Class IITerminatedVoluntary: Firm initiated