DePuy Synthes various orthopedic instruments modified by U.S. Distributors These instruments are used in...
DePuy Orthopaedics, Inc. began a recall of DePuy Synthes various orthopedic instruments modified by U.S. Distributors These instruments are used in various orthopedic procedures on . FDA classified it as Class II and gives the reason as "Products were made outside of Quality System Regulation, and potentially outside of premarket submission (510k/PMA) for certain devices. The safety or effectiveness of these devices cannot be assured". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-2317-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Depuy Orthopaedics Inc.
- Product
- DePuy Synthes various orthopedic instruments modified by U.S. Distributors These instruments are used in various orthopedic procedures
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Products were made outside of Quality System Regulation, and potentially outside of premarket submission (510k/PMA) for certain devices. The safety or effectiveness of these devices cannot be assured.
- Root cause tag
- supplier
- FDA root cause
- Vendor change control
- Action
- DePuy Synthes sent an URGENT INFORMATION RECALL NOTICE FOR INSTRUMENTS MODIFIED BY U.S. DISTRIBUTORS dated January 6, 2017, to all affected customers. . Customerss were instructed to identify all medical facilities that may have used or received the affected instruments and identify the modified instruments used at each facility. This information was to be used to generate Reconciliation Forms for each impacted Medical Facility. The Reconciliation Forms and URGENT INFORMATION RECALL NOTICE FOR INSTRUMENTS MODIFIED BY U.S. DISTRIBUTORS recall notifications were then delivered by DePuy Synthes Sales Consultants to the affected medical facilities. Instructions in the URGENT INFORMATION RECALL NOTICE FOR INSTRUMENTS MODIFIED BY U.S. DISTRIBUTORS provided to the medical facilities included the following: Please take the following actions: " Please immediately cease using the modified instruments identified in the attached Reconciliation Form. Your U.S. DePuy Synthes Sales consultant will work with your facility to locate and replace any affected instruments. " If your facility is using an instrument that was created or modified by a DePuy Distributor at the request of a surgeon, and it is not listed on the attached Reconciliation Form, please contact your DePuy Synthes Sales Consultant for an evaluation to determine if the instrument should be returned and replaced. WI-9956 | Rev 5 | Attachment B2 " Return Affected Instruments: o Medical facilities are to determine if any of the recalled instruments are still on hand by working with your U.S. DePuy Synthes Sales Consultant, and return affected devices immediately to their U.S. DePuy Synthes Sales Consultant or return them to DePuy Orthopaedics, Inc. for credit following normal purchasing procedures. o Note: These instruments may be on consignment at your facility. " Reconciliation Form: Complete the Reconciliation Form and return to your U.S. DePuy Synthes Sales Consultant or
- Quantity in commerce
- 345
- Distribution
- Nationwide Distribution to: AZ CA IA IL IN LA MA MD ME MI MN PA VA
- Product codes and lots
- Unknown
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2317-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.