MagSIL (NucliSENS easyMAG Magnetic Silica)
BioMerieux SA began a recall of MagSIL (NucliSENS easyMAG Magnetic Silica), Product Usage: The NucliSENS easyMAG system is an in vitro diagnostic medical device and is intended for the... on . FDA classified it as Class I and gives the reason as "A drift of the performance with the BK test (BK virus, worst case application) was observed for some silica batches of Magnetic Silica (MagSil). This is the same issue as in recall Z-2342-2016". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0976-2017
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- bioMerieux, Inc.
- Product
- MagSIL (NucliSENS easyMAG Magnetic Silica), Product Usage: The NucliSENS easyMAG system is an in vitro diagnostic medical device and is intended for the automated isolation (purification and concentration) of total nucleic acids (RNA/DNA) from biological specimens.
- Product code
- PPM General Purpose Reagent
- Reason for recall
- A drift of the performance with the BK test (BK virus, worst case application) was observed for some silica batches of Magnetic Silica (MagSil). This is the same issue as in recall Z-2342-2016.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Consignees were notified via, URGENT PRODUCT SAFETY CORRECTION NOTICE on/about 11/23/2016. This is the same issue as in recall Z-2342-2016, notification letter sent on/about 7/8/2016.
- Quantity in commerce
- 3723 units in total
- Distribution
- Worldwide distribution. US Nationwide; BioMerieux Subsidiaries in: Argentina, Austria, Australia, Brazil, Canada, Switzerland, Chile, Colombia, Czech republic, Germany, Spain, France United Kingdom, Greece, China, Hungary, India, Italy, Japan, South Korea, Netherlands, Poland, Portugal, Russia Fed., Sweden, Singapore, Thailand, Turkey, South Africa; Distributors in: Peru, Kenya, Morocco, Serbia, Romania, Uruguay, Latvia, Dutch Antilles, Israel, Georgia, Armenia, Malaysia, Belarus and Direct Customers in: Togo, Djibouti, Burkina Faso, Mongolia, Zimbabwe, Uganda and Iraq.
- Product codes and lots
- Reference number: 280133, Lot numbers: Z017KA1MS, Z017KB1MS, Z017LE1MS, Z017MA1MS, Z017NA1MS, Z017NB1MS, Z017NC1MS and Z017ND1MS.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0976-2017 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0977-2017 | Class I | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
