Z-1026-2017Class II

Fresenius 2008T Series Hemodialysis System

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2016-12-21 / Terminated / root cause: design

Fresenius Medical Care Renal Therapies Group, LLC began a recall of Fresenius 2008T Series Hemodialysis System on . FDA classified it as Class II and gives the reason as "When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1026-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
Fresenius 2008T Series Hemodialysis System
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine
Root cause tag
design
FDA root cause
Device Design
Action
Fresenius issued Customer Notification via Certified Mail on 12/21/2016. Users provided with problem and recommendation, when adjusting the UF Rate, UF Goal, or Tx Time, is to use the numeric data entry keys instead of the up or down arrow keys. Change the values by using data entry keys, then save by pressing CONFIRM. Fresenius will make arrangements to install software on the hemodialysis machine. Confirmation of the noticed by the consignee will be required via a Reply Form.
Quantity in commerce
65361 units (US) 249 units (OUS)
Distribution
Nationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam
Product codes and lots
All serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1026-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1026-2017Class IITerminatedVoluntary: Firm initiated
Z-1027-2017Class IITerminatedVoluntary: Firm initiated
Z-1028-2017Class IITerminatedVoluntary: Firm initiated
Z-1029-2017Class IITerminatedVoluntary: Firm initiated