Fresenius 2008K Series : Hemodialysis System
Fresenius Medical Care Renal Therapies Group, LLC began a recall of Fresenius 2008K Series : Hemodialysis System on . FDA classified it as Class II and gives the reason as "When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1027-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Product
- Fresenius 2008K Series : Hemodialysis System
- Product code
- KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
- Reason for recall
- When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Fresenius issued Customer Notification via Certified Mail on 12/21/2016. Users provided with problem and recommendation, when adjusting the UF Rate, UF Goal, or Tx Time, is to use the numeric data entry keys instead of the up or down arrow keys. Change the values by using data entry keys, then save by pressing CONFIRM. Fresenius will make arrangements to install software on the hemodialysis machine. Confirmation of the noticed by the consignee will be required via a Reply Form.
- Quantity in commerce
- 78481 (US) 2734 units (OUS)
- Distribution
- Nationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam
- Product codes and lots
- All serial numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1026-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1027-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1028-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1029-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.