Z-1098-2017Class II
Softbank software
SCC Soft Computer / initiated 2016-10-06 / Terminated / root cause: software
Soft Computer Consultants, Inc. began a recall of Softbank software Product Usage: Decision support software for transfusion service. It keeps track of inventory from outside sources, multi-site inventory... on . FDA classified it as Class II and gives the reason as "Software error. Potential for incorrect results". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1098-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- SCC Soft Computer
- Product
- Softbank software Product Usage: Decision support software for transfusion service. It keeps track of inventory from outside sources, multi-site inventory control, records of testing of units, and allows for record keeping for transfusion preparation.
- Product code
- NSX
- Reason for recall
- Software error. Potential for incorrect results
- Root cause tag
- software
- FDA root cause
- Software design (manufacturing process)
- Action
- Consignees were sent a recall notification, dated 10/06/2016, providing instructions for software correction. The firm indicates that the issue will be resolved with a mandatory hot fix. Hot fix will be released when its effectiveness is confirmed. Consignees are instructed to acknowledge receipt of the recall notification and grant permission for the recalling firm to load the hot fix when available. The firm can be contacted by calling Technical Support at 1-800-763-8522. Until the hot fix is available, consignees can still operate the software by exiting (Esc-Quit) the Test Verify option after each save changes (F12-Accept).
- Quantity in commerce
- 189 total
- Distribution
- Worldwide Distribution - US Nationwide Distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WA, WI, WV, and the countries of Hawaii and Puerto Rico Canada & Jamaica
- Product codes and lots
- Version: 35.5.2.0, 25.5.3.0, 25.5.3.1
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1097-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1098-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY