Z-1096-2017Class II
Hitachi Oasis MRI System - C-Spine Coil
FUJIFILM Healthcare Americas Corporation / initiated 2017-01-06 / Terminated / root cause: design
Hitachi Medical Systems America Inc began a recall of Hitachi Oasis MRI System - C-Spine Coil on . FDA classified it as Class II and gives the reason as "The customer indicated that the coil balun was hot to the touch when removing the coil from the patient table after an exam". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1096-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Product
- Hitachi Oasis MRI System - C-Spine Coil
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- The customer indicated that the coil balun was hot to the touch when removing the coil from the patient table after an exam.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Hitachi sent an Urgent : Device Correction Notice dated January 3, 2017, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm plans to visit each site and replace the cable balun and also confirm receipt of the notification letter. The firm will quarantine returned baluns in their warehouse until they are reworked. For further questions regarding this recall please call (330) 425-1313 ext. 3720.
- Quantity in commerce
- 180 units
- Distribution
- USA (nationwide) distribution. The product was distributed to the following states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NH, NJ, NM, NY, OH, OK, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY. There are three (3) government accounts for this recall: VA Medical Center/Lyster Army Health Clinic, Parma VA Medical Center, Outpatient Clinic. There are no Canada accounts for this recall. There are no Mexico accounts for this recall.
- Product codes and lots
- M001 to M028, M029 to M083, M085 to M090, M092 to M104, M106, M110 to M111, M113 to M114, M116 to M123, M125, M128 to M131, M133, M139 to M141, M143 to M147, M150, M152, M154, M159, M161, M163, M166, M169 to M176, M179, M182, M183, M185, M186, M190, M194 to 196, M201, M207 to M238
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1096-2017 | Class II | Terminated | Voluntary: Firm initiated |