Z-1007-2017Class II

AxIEM Touch N Go Pointer Probe. Model Number: 9735318. Placed in a 30mil PETG tray sealed with a Tyvek lid...

Medtronic Navigation, Inc / initiated 2016-12-12 / Terminated

Medtronic Navigation, Inc. began a recall of AxIEM Touch N Go Pointer Probe. Model Number: 9735318. Placed in a 30mil PETG tray sealed with a Tyvek lid and then placed in a shelf carton on . FDA classified it as Class II and gives the reason as "Medtronic Navigation, Inc. announces a voluntary field action for the AxIEM Touch N Go Pointer Probe because the patient registration accuracy may compromise navigation accuracy when using the StealthStation System with...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1007-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
AxIEM Touch N Go Pointer Probe. Model Number: 9735318. Placed in a 30mil PETG tray sealed with a Tyvek lid and then placed in a shelf carton.
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Medtronic Navigation, Inc. announces a voluntary field action for the AxIEM Touch N Go Pointer Probe because the patient registration accuracy may compromise navigation accuracy when using the StealthStation System with Touch N Go Pointer Probes for patient registration.
FDA root cause
Process change control
Action
The firm, Medtronic, sent an "URGENT: MEDICAL DEVICE RECALL" letter on 12/12/16 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to examine inventory for any affected products, quarantine them for return to Medtronic; complete and return the CUSTOMER RESPONSE FORM by email at [email protected] or fax to Medtronic Technical Services at 651-367-7075 (even if you do not have affected product); and contact Medtronic Technical Services at 1-800-595-9709 to get a return material authorization (RMA) and arrange for their return and no charge replacements. If you have any questions regarding this communication, please contact Medtronic Technical Services at (800) 595-9709.
Quantity in commerce
111 devices
Distribution
US Nationwide Distribution.
Product codes and lots
lot number 150626B

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1007-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1007-2017Class IITerminatedVoluntary: Firm initiated