Z-1060-2017Class II

OEC 9900 Elite. MDL Numbers: D148942, D155043

GE Healthcare / initiated 2016-12-27 / Terminated / root cause: packaging

GE OEC Medical Systems, Inc began a recall of OEC 9900 Elite. MDL Numbers: D148942, D155043 on . FDA classified it as Class II and gives the reason as "GE Healthcare Surgery announces a voluntary field action for the OEC 9800 and OEC 9900 Elite due to additional supplementation to the user documentation to include areas for enhanced planned maintenance and component...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1060-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
OEC 9900 Elite. MDL Numbers: D148942, D155043
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
GE Healthcare Surgery announces a voluntary field action for the OEC 9800 and OEC 9900 Elite due to additional supplementation to the user documentation to include areas for enhanced planned maintenance and component replacement frequencies.
Root cause tag
packaging
FDA root cause
Packaging change control
Action
The customer communication is expected to begin distribution in January 2017. Consignees will be notified of the availability of the Operator Manual supplement via a letter. No product is being returned.
Quantity in commerce
26,256 total
Distribution
Worldwide including govt/VA/military and foreign consignees.
Product codes and lots
Manufactured from Oct 2005 to present

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1060-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1059-2017Class IITerminatedVoluntary: Firm initiated
Z-1060-2017Class IITerminatedVoluntary: Firm initiated