Z-1061-2017Class II
Velara Generator with an old Firmware version used with Philips Allura Xper Systems (Allura 15 & 12...
Philips North America LLC / initiated 2016-12-09 / Terminated / root cause: design
Philips Electronics North America Corporation began a recall of Velara Generator with an old Firmware version used with Philips Allura Xper Systems (Allura 15 & 12 (monoplane), Allura FD10C, Allura FD10F, Allura FD 10 F/C... on . FDA classified it as Class II and gives the reason as "System may lock up". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1061-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Velara Generator with an old Firmware version used with Philips Allura Xper Systems (Allura 15 & 12 (monoplane), Allura FD10C, Allura FD10F, Allura FD 10 F/C, Allura FD20, Allura XPER FD10/10, Allura XPER FD20 Biplane)
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- System may lock up.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Consignees were notified of the recall by letter, sent via certified mail on 12/9/16. Philips will upgrade the firmware of the Velara Generator, allowing the Velara Generator to recover in case of intermittent shortcuts. This service will be provided free of charge for all affected systems and will start effective January 2017. A Philips service representative will contact customers with affected devices to arrange for the service.
- Quantity in commerce
- 59 devices (this number represents all products listed in recall)
- Distribution
- US: AL, CA, DC, IL,GA, KS, MD, MO, NC, NY, OH, PA, RI, SC, TX, WA OUS: Brazil, Bulgaria, Costa Rica, France, Germany, Israel, Italy, Japan, Korea, Republic of, Mexico, Russian Federation, Singapore, Spain, Sweden, Switzerland Switzerland, Turkey, United Kingdom
- Product codes and lots
- Velara Generator with 12nc number: 989000070011, 989000070122, 989000070123, 989000070124, 989000070125, 989000070126, 989000070321
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1061-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1062-2017 | Class II | Terminated | Voluntary: Firm initiated |