SwishTapered Implant
Implant Direct Sybron Manufacturing, LLC began a recall of SwishTapered Implant on . FDA classified it as Class II and gives the reason as "The extender product code (9037-03) that is packaged with the SwishTapered (Part Number 933308; Lot Number 73241) Implant was packaged in the incorrect position. This discrepancy may lead to the possibility of extender...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1112-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Implant Direct Sybron Manufacturing LLC
- Product
- SwishTapered Implant
- Product code
- DZE Implant, Endosseous, Root-Form
- Reason for recall
- The extender product code (9037-03) that is packaged with the SwishTapered (Part Number 933308; Lot Number 73241) Implant was packaged in the incorrect position. This discrepancy may lead to the possibility of extender not seating properly on the implant, which may lead to micro-leakage and a possible infection to the patient.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Customer notification letters were sent on 12/12/16. Customers were requested to review inventory for the affected product. Complete and return the Acknowledgement and Recall Return Form within 48 hours. If customers are an authorized distributor the firm requests that they identify those customers that may have been shipped the affected product lot and contact those customers to inform them of the issue within 48 hours of receipt of the notification in order to provide the customers with replacement product. If customers have any of the affected product the firm will send a replacement part. Any questions contact Implant Direct Sybron Manufacturing LLC Customer Care at 1-888-649-6425.
- Quantity in commerce
- 17
- Distribution
- U.S. distribution to the following; NM, TX, CA, OR, OH, ME. Foreign distribution to the following; Netherlands.
- Product codes and lots
- 933308 Lot # 73241
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1112-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY