Z-1117-2017Class II

VITROS Na+ Slides, (UPN 10758750004812) For in vitro diagnostic use only. VITROS Na+ Slides quantitatively...

Ortho-Clinical Diagnostics, Inc. / initiated 2016-12-20 / Terminated / root cause: labeling and UDI

Ortho-Clinical Diagnostics began a recall of VITROS Na+ Slides, (UPN 10758750004812) For in vitro diagnostic use only. VITROS Na+ Slides quantitatively measure sodium (Na+) concentration in serum, plasma... on . FDA classified it as Class II and gives the reason as "There is a potential for biased results to be generated over the 10 day on-analyzer storage limit when using VITROS Na+ Slide cartridges warmed at room temperature and loaded onto the analyzer after using the minimum...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1117-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Na+ Slides, (UPN 10758750004812) For in vitro diagnostic use only. VITROS Na+ Slides quantitatively measure sodium (Na+) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Product code
JGS Electrode, Ion Specific, Sodium
Reason for recall
There is a potential for biased results to be generated over the 10 day on-analyzer storage limit when using VITROS Na+ Slide cartridges warmed at room temperature and loaded onto the analyzer after using the minimum warm up requirements stated in the IFU. If Quality Control (QC) results are within acceptable limits, reported results are acceptable and were not affected by this issue.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
Ortho Clinical Diagnostics sent a Urgent Product Correction Notification dated December 20, 2016, to all affected customers via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all customers who have a VITROS analyzer that can process Na+ Slides to inform them of the potential for biased results on Na+ Slides over the 10 day on-analyzer storage limit for Na+ Slides. Customers were instructed to warm up (equilibrate) all Na Slide cartridges for a minimum of eight hours at room temperature prior to loading them onto the analyzer to help ensure the slides have acceptable performance throughout the 10 day on-analyzer storage limit. Foreign affiliates were informed by email on December 20, 2016, of the issue and instructed to notify their Customers of the issue and the appropriate actions to take. For further questions, please call (585) 453-4224.
Quantity in commerce
659,957 Total Units (249,570 units domestically & 410,387 units internationally)
Distribution
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain and Poland.
Product codes and lots
Generation (GEN): 1 through 6, 30 through 40, 44 through 49 Expiry Dates: 01-JAN-2017 through 01-JUN-2018

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1117-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1117-2017Class IITerminatedVoluntary: Firm initiated