Unna-Z, Unna Boot with Zinc and Calamine, Compression Wrap, 1 Per Box, 3-4in x 10yds
Medline Industries Inc began a recall of Unna-Z, Unna Boot with Zinc and Calamine, Compression Wrap, 1 Per Box, 3-4in x 10yds on . FDA classified it as Class II and gives the reason as "Medline Industries, Inc is recalling Unna Z Unna Boot Bandage 3X10 and Unna Z Unna Boot Bandage 4X10 due to contamination of the products with common mold (Cladosporium...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1074-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Unna-Z, Unna Boot with Zinc and Calamine, Compression Wrap, 1 Per Box, 3-4in x 10yds
- Product code
- FQM Bandage, Elastic
- Reason for recall
- Medline Industries, Inc is recalling Unna Z Unna Boot Bandage 3X10 and Unna Z Unna Boot Bandage 4X10 due to contamination of the products with common mold (Cladosporium sphaerospermum/cladosporioides/herbarum/phaenocomae/halotolerans and Penicillium corylophilum)..
- Root cause tag
- manufacturing process
- FDA root cause
- Environmental control
- Action
- Medline Industries sent an Immediate Action letter dated November 23, 2016, to all affected customers with response forms to their customers. Customers were instructed to destroy the affected Unna Boot Bandages. Affected lots of product remaining at the firm's distribution centers have been placed on hold and quarantined pending destruction. All affected product will be destroyed. For further questions, please call (847) 643-3245.
- Quantity in commerce
- 117,059 units
- Distribution
- Worldwide Distribution - USA (nationwide) and to the countries of : Chile, Peru, El Salvador and Taiwan.
- Product codes and lots
- Unit Numbers: NONUNNA3 & NONUNNA4 Lot Numbers: 52581 through 63599