Z-1058-2017Class II

Brilliance iCT

Philips North America LLC / initiated 2016-12-23 / Terminated / root cause: manufacturing process

Philips Medical Systems (Cleveland) Inc began a recall of Brilliance iCT on . FDA classified it as Class II and gives the reason as "During a retrospective review of production documentation it was observed that the M6 Rotor Counterweight Bolt may have been improperly torqued". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1058-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Brilliance iCT
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
During a retrospective review of production documentation it was observed that the M6 Rotor Counterweight Bolt may have been improperly torqued.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Philips sent an Customer Information letter. A document showed an incorrect torque wrench used during the assembly process which could potentially apply 4 Newton-meters less torque than specified to the system rotor counterweight bolts. Out of an abundance of caution, Philips will have a Field Service Engineer ensure the bolts on your system are correctly torqued. If you need any further information or support concerning this issue, please contact your local Philips representative. For further information please call (440) 483-5777.
Quantity in commerce
1-unit
Distribution
Product was not distributed in the US. Unit was distributed in Australia.
Product codes and lots
Rotor-4535-673-94382 SN 19, installed on Brilliance ICT System S/N 200203

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1058-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1058-2017Class IITerminatedVoluntary: Firm initiated