Z-1058-2017Class II
Brilliance iCT
Philips North America LLC / initiated 2016-12-23 / Terminated / root cause: manufacturing process
Philips Medical Systems (Cleveland) Inc began a recall of Brilliance iCT on . FDA classified it as Class II and gives the reason as "During a retrospective review of production documentation it was observed that the M6 Rotor Counterweight Bolt may have been improperly torqued". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1058-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Brilliance iCT
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- During a retrospective review of production documentation it was observed that the M6 Rotor Counterweight Bolt may have been improperly torqued.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Philips sent an Customer Information letter. A document showed an incorrect torque wrench used during the assembly process which could potentially apply 4 Newton-meters less torque than specified to the system rotor counterweight bolts. Out of an abundance of caution, Philips will have a Field Service Engineer ensure the bolts on your system are correctly torqued. If you need any further information or support concerning this issue, please contact your local Philips representative. For further information please call (440) 483-5777.
- Quantity in commerce
- 1-unit
- Distribution
- Product was not distributed in the US. Unit was distributed in Australia.
- Product codes and lots
- Rotor-4535-673-94382 SN 19, installed on Brilliance ICT System S/N 200203
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1058-2017 | Class II | Terminated | Voluntary: Firm initiated |