Z-1141-2017Class II

Dimension Vista Ammonia Flex reagent cartridge / AMM Dimension: The AMM method is an in vitro diagnostic test...

Siemens Healthcare Diagnostics Inc. / initiated 2016-12-16 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista Ammonia Flex reagent cartridge / AMM Dimension: The AMM method is an in vitro diagnostic test for the quantitative measurement of ammonia in... on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics has determined that Dimension Vista AMM (K3119) Flex reagent cartridge flex lots 16187BE, 16225BB, 16265AB do not meet the 60-day calibration interval claim due to reagent instability and...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1141-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension Vista Ammonia Flex reagent cartridge / AMM Dimension: The AMM method is an in vitro diagnostic test for the quantitative measurement of ammonia in human plasma on the Dimension clinical chemistry system. Ammonia measurements are used in the diagnosis and treatment of severe liver disorders such as cirrhosis, hepatitis and Reye s syndrome
Product code
JIF Enzymatic Method, Ammonia
Reason for recall
Siemens Healthcare Diagnostics has determined that Dimension Vista AMM (K3119) Flex reagent cartridge flex lots 16187BE, 16225BB, 16265AB do not meet the 60-day calibration interval claim due to reagent instability and results may show an Abnormal Assay.
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
Siemens sent an Urgent Field Safety Notice on December 20, 2016, to all customers who were affected by the issue. The letter identified the reason for the correction, the associated risks to health, and actions to be taken by the customer. Customers were instructed to discontinue use of and discard the affected lots, complete and return the Field Correction Effectiveness Check Form attached to the letter. Customers with questions were instructed to contact their Siemens Customer Care Center or their local Siemens technical support representative. For questions regarding this recall call 800441-9250.
Quantity in commerce
8,032 units
Distribution
Worldwide Distribution - US (Nationwide) and Internationally to Argentina, Austria, Australia, Bangladesh, Belgium, Brazil, Bosnia Herzegovina. ,Canada, Canary Islands, Chile, China, Colombia, Croatia, Denmark, Ecuador, Egypt, France, Germany, Great Britain, Hong Kong, Hungary Israel ,India, Indonesia, Italy, Lebanon, Malaysia, Mexico, Nicaragua, Netherland, New Zealand, Philippines, Poland, Portugal, Republic of Korea, Saudi Arabia, Serbia, Slovakia, Slovenia, Spain, Sri Lanka, South Africa, Switzerland, Taiwan, Turkey, Uruguay, United.Arab Emirates, Venezuela, Yemen, and Japan.
Product codes and lots
lots 16187BE, 16225BB and 16265AB

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1141-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1140-2017Class IITerminatedVoluntary: Firm initiated
Z-1141-2017Class IITerminatedVoluntary: Firm initiated