Z-1191-2017Class II

LIMA Modular Revision Hip Stem Model 428-01-050_110

Encore Medical, LP / initiated 2017-01-10 / Terminated / root cause: manufacturing process

Encore Medical, Lp began a recall of LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a proper... on . FDA classified it as Class II and gives the reason as "Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1191-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encore Medical, LP
Product
LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a proper neck by means of a Morse taper stabilized during the implantation phase by a safety screw. This system is particularly indicated for revision surgery on both uncemented and cemented femoral implants, when there is significant bone loss and an abnormal meta-epiphyseal anatomy of the femur.
Product code
KWA Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)
Reason for recall
Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
DJO Global sent an Urgent Field Safety Notice letter dated January 10, 2017 to customers. The letter identified the affected product, problem and actions to be taken. Customers were asked to complete the Acknowledgement and receipt Form. Customers were instructed to contact Customer Service at 1-800-456-8696 to explore options for a replacement order.
Quantity in commerce
13 devices
Distribution
US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI
Product codes and lots
1005528A, 1006882A, 1107305A, 1101798A, 1100565A, 1005531A, 1211716A, 1007408A, 1004038B, 1005533D, 1106805A, 1203471A, 1005533C, 1100565A, 1101795A, 1007079A, 1005532A, 1100560A

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1191-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1191-2017Class IITerminatedVoluntary: Firm initiated