Z-1228-2017Class II
Vascular catheter introduction kit The PICC is indicated for short or long term peripheral access to the...
Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2017-01-11 / Terminated / root cause: design
Arrow International Inc began a recall of Vascular catheter introduction kit The PICC is indicated for short or long term peripheral access to the central Venous system for intravenous therapy, blood... on . FDA classified it as Class II and gives the reason as "There have been complaints for peel away sheaths flaring". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1228-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arrow International LLC (subsidiary of Teleflex Incorporated)
- Product
- Vascular catheter introduction kit The PICC is indicated for short or long term peripheral access to the central Venous system for intravenous therapy, blood sampling, infusion and power injection of contrast media.
- Product code
- LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days
- Reason for recall
- There have been complaints for peel away sheaths flaring
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Arrow International mailed an Urgent Medical Device Advisory Notification letter to affected customers on 01/13/2017 to inform them of the issue. Customers with questions were instructed to contact their local sales representative or Customer Service at 1-866-246-6990.
- Quantity in commerce
- 215,703 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- Material # ASK-01663-MST, ASK-04001-DU4, ASK-04001-DU5, PI-01351-LS, PI-01351-LS5, PI-01351-SS, PI-01552-LS, PI-01552-LS5, PI-01552-SS, PL-05041, PL-05041, PL-05052, PR-05041, PR-05041, PR-05041-T, PR-05042, ASK-01663-MST, ASK-04001-DU4, ASK-04001-DU5, PI-01351-LS, PI-01351-LS5, PI-01351-SS, PI-01351-SS, PI-01451-LS, PI-01451-LS5, PI-01451-SS, PI-01451-SS, PI-0-1552-LS
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1228-2017 | Class II | Terminated | Voluntary: Firm initiated |